This test measures the Voltage Standing Wave Ratio (VSWR) of the Keysight power sensor. A network analyzer accompanied with a calibration kit or electronic calibration modules is used to perform this test.
Instrument | Recommended product | Alternative product | U2000A,U2000B, U2000H | U2001A,U2001B, U2001H | U2002A,U2002H
|
U2004A
|
Network analyzers | ||||||
PNA1 |
N5247B |
N5245A/AS E8361C |
X | X | X | |
ENA | E5071C3 | - | X | |||
Calibration Kits/Electronic Calibration Modules | ||||||
50 Ω Calibration Kit | 85054B |
N4690B Opt M0F or 85032F (for U2001A, U2001B, and U2001H only) |
X | X | ||
50 Ω Calibration Kit | 85032F | - | X | |||
50 Ω Calibration Kit | 85052B | N4691B Opt M0F or N4691BH33 Opt M0F |
X | |||
Meters | ||||||
Power Meter4 | N1914A | N1912A E4419B |
X | |||
Power Sensor5 | N8481A |
N8481B or N8481H or 8481A or 8481B or 8481H |
X | X | ||
Power Sensor5 | N8485A |
8485A | X | |||
Power Sensor5 | E9304A | - |
X | |||
Adapters6 | ||||||
2.4 mm (f) to 3.5 mm (f) | 11901B |
- | X | |||
2.4 mm (f) to Type-N (f) | 11903B |
- | X | X |
|
An E-Cal may be used for the network analyzer calibration instead of a mechanical calibration kit. When using the E-Cal, please connect it to the network analyzer using a USB cable. Before using the E-Cal, ensure that the ready light has turned green on the unit. When mapping the E-Cal in TME, the serial number should be a 5 digit number with no letters. If there is no 5 digit serial number printed on your unit, then follow these instructions using your network analyzer to locate your serial number:
Using Calibration Kit:
Using Electronic Calibration Modules:
Using N8480 Series or 8480 Series power sensor:
Using E9304A power sensor:
For all UUT, except U2000B and U2001B:
For U2000B and U2001B: