Downconverter RF Attenuator Adjustment

(M9391A)

This adjustment varies RF attenuators in all RF paths, over frequency range, to minimize any errors.

During this adjustment, the lights of the PXIe signal generator will not be on even though it is running.

Required Test Equipment

Test Equipment1

Recommended Models2

Signal generator

M9381A Opt 1EA, F06

Fixed attenuator, 6 dB

8493C Opt 006

Frequency reference module

M9300A

  1. Refer to the main required equipment list for a list of available adapters and cables.
  2. Refer to the main required equipment list for alternative models.

Connections and Setup Procedures