Adjustments Overview

If any performance test should fail, a technician will be required to run all adjustments.

  • Running an adjustment will invalidate the calibration date of that module as well as the calibration information for all PXIe systems that the module belongs to.

  • The M9300A frequency reference module must be installed in slot 10 of the chassis.

  • In all test equipment configurations, cables and adapters must be properly torqued. Exceeding recommended torque values can cause damage to a cable or adapter, and can cause inaccurate test results.

Click on the links below to go to the adjustment topic page.

  1. Synthesizer VCO

  2. Synthesizer Output Power

  3. Synthesizer Output Rel Power