Frequency Response Gain and Output Power

(M9340A)

This test measures the frequency response gain, minimum and maximum output power of the M9340A distributor module. The module consists of a local and source signal block. Each signal block contains a through path, prescaler, low frequency doubler and high frequency doubler. Frequency response gain is measured using the through path, while output power is measured using the prescaler and low/high frequency doubler paths. A network analyzer is used to measure the gain and output power.

This test requires that a USB-to-GPIB adapter and an E-Cal module be connected to the E5071C network analyzer. A power meter will be connected to the USB-GPIB adapter to allow the E5071C network analyzer to control the device.

Required Test Equipment

Test Equipment

Recommended Model Number1

Network analyzer

E5071C Option 4K5 or 4D5

Power meter

N1914A

Power sensor

E9304A-H18

ECal module

85093-60010

USB-to-GPIB adapter

82357B

  1. Refer to the main equipment list for alternative models.

Connection Setups

To expand all, click the Expand All button in the toolbar (webhelp only).

Calibrate the source. ClosedClick here.

 

Making the measurement. ClosedClick here.