N7878A Required Test Equipment

This section lists the test equipment needed to run performance tests. Models for each test equipment category are given in order of preference. Alternate models are also listed but should be considered a lower priority. All equipment is Keysight, Agilent, or Hewlett Packard unless otherwise specified. Refer to each individual test for specific setup configurations. The computer and all test equipment are connected to the unit under test (UUT) using GPIB and PXIe.

Test Equipment

Recommended Model

Alternate Model

Network analyzer

E5071C1 Option 4K52 or 4D52


Power meter

N1914A3

N1912A
E4419A/B

Power sensor
(See below for calibration service ordering guidelines.)

For M9376A: E9304A-H18

E9304A-H19

For M9377A: E9304A-H19

E9304A-H184

For M9340A: E9304A-H18

none

Power sensor
(See below for calibration service ordering guidelines.)

N8482A

8482A


Dynamic accuracy test set

Z5623A-H01

 

Calibration kit, 3.5 mm

85052D, mechanical
(M9376A/77A)

85093-60010 (85093C-00F), ECal
(M9377A only)

ECal module

85093-60010 (85093C-00F)
(M9340A, M9378A/B)

 

6 dB fixed attenuator

8491B Opt. 006

8491A Opt. 006

10 dB step attenuator6, 7

8496G Option H525

8496H Option H525

Attenuator switch driver

11713B

11713A

50 Ω load
(2 required)

Any
(can come from the 85052D cal kit)

 

Adapter for receiver module

M9376-65001

 

USB-to-GPIB adapter

82357B

 

Table Footnotes:

  1. The frequency of the E5071C network analyzer should be greater than or equal to the maximum frequency of the UUT.

  2. Option 4K5 or 4D5 required for the M9340A distributor and M9378A/B directional coupler modules.

  3. N1914A power meters with serial number prior to MY53040007 require Service Note N1914A-07 which fixes a power supply ground loop. The ground loop injects noise into the measurement circuits which can cause unstable measurements at low (-30 dBm) power levels.

  4. The E9304A-H18 power sensor is not allowed for the M9377A direct access receiver Maximum Leveled Power test since the power measured will exceed the power supported by the E9304A-H18 sensor.

  5. Option H52 is required for the M9376A receiver.

  6. The 10 dB step attenuator should be a standalone step attenuator. It should not be joined to a 1 dB step attenuator.

  7. Step Attenuator Calibration

    Please have the 10 dB step attenuator calibrated at 1 MHz for the M9376A receiver or at 50 MHz for the M9377A direct access receiver for all steps, and enter the calibration data in TME prior to running any tests. The left port of the attenuator will be used for testing and for entering calibration data in TME.

    The following tables show which sections of the 10 dB step attenuator are utilized for each attenuator setting. The tables also list the Recommended Uncertainty for each attenuator setting. A larger number will result in larger overall test uncertainties which could affect the test pass/fail rate.

    10 dB Step Attenuator

    Nominal Attenuation

    (dB)

    Attenuator Section

    Recommended

    Uncertainty

    (dB)

    #1

    (10 dB)

    #2

    (20 dB)

    #3

    (40 dB)

    #4

    (40 dB)

    0a

    Off

    Off

    Off

    Off

    0 (Reference)

    10

    On

    Off

    Off

    Off

    < 0.010

    20

    Off

    On

    Off

    Off

    < 0.015

    30

    On

    On

    Off

    Off

    < 0.020

    40

    Off

    Off

    Onb

    Off

    < 0.025

    50

    On

    Off

    On

    Off

    < 0.030

    60

    Off

    On

    On

    Off

    < 0.035

    70

    On

    On

    On

    Off

    < 0.040

    1. TME requires that there be 0 dB of Uncertainty and 0 dB of Attenuation for the 0 dB reference step.
    2. TME requires that Section 3 be characterized.

    Example of 10 dB Step Attenuator Calibration Data

    Frequency

    Step

    Attenuation

    Reflection
    Sizea

    Uncertainty

    Reflection
    Anglea

    M9376A

    M9377A

    1000000

    50000000

    0

    0b

    0.009

    0b

    –4.58

    1000000

    50000000

    1

    0.996

    0.009

    0.006

    –4.93

    1000000

    50000000

    2

    1.998

    0.01

    0.006

    –12.89

    1000000

    50000000

    3

    2.994

    0.009

    0.006

    –11.77

    1000000

    50000000

    4

    4.017

    0.007

    0.006

    33.15

    1000000

    50000000

    5

    5.014

    0.007

    0.01

    25.06

    1000000

    50000000

    6

    6.017

    0.007

    0.009

    29.72

    1000000

    50000000

    7

    7.013

    0.007

    0.008

    22.1

    1000000

    50000000

    8

    8.051

    0.007

    0.007

    22.96

    1000000

    50000000

    9

    9.046

    0.007

    0.009

    16.56

    1000000

    50000000

    10

    10.051

    0.007

    0.008

    23.04

    1000000

    50000000

    11

    11.047

    0.007

    0.008

    16.53

    1. This calibration data is not used in N7878A calibration application, thus default value (zero) entry is allowed.
    2. TME requires that there be 0 dB of Uncertainty and 0 dB of Attenuation for the 0 dB reference step.

Guidelines for Ordering Power-Sensor Calibration Service

Use the following guidelines to ensure power sensors used by the N7800A application receive the correct calibration service:

  1. Order Option 1A7 when ordering new power sensors that will be used as working standards in N7800A TME calibrations.

  2. In the Americas region, order the Keysight Technologies Cal + Measurement Uncertainty for power sensor recalibration.

  3. Outside the Americas region, order ISO 17025/ILAC-G8 calibration service for power sensor recalibration.

  4. When ordering periodic calibration for instruments used as lab standards in the N7800A software, we recommend using “Keysight Technologies calibration + uncertainties” for power sensors, and “Keysight Technologies calibration + uncertainties + guardbanding” for all other items (please visit Selecting the Right Calibration Services). The N7800A software incorporates the ISO GUM Uncertainty in point-to-point uncertainty calculations. The special “H-series” calibration options in this table provide lower measurement uncertainties through use of direct comparison to devices directly characterized by NPL or NIST (or another NMI). Please order Option H99 to get data on a CD for easy import into the N7800A (avoids manual entry). The overall resulting N7800A measurement uncertainties then reflects these lower device uncertainties. The equipment requirements of each N7800A calibration application are summarized in the specials calibration matrix which can be found at the Recommended Lab Standards and Special Cal Options website.

  5. The Keysight Technologies Option H99 is a special option for the Roseville Service Center [only] which provides a CD with calibration data included as a .csv file in TME N7800A format for all power sensors used with the N7800A TME application. N7800A TME is SSU calibration software used across Keysight Technologies SSU and by many self-maintainers. H99 must be requested from the Roseville SSU upon re-calibration only of any previously purchased power sensor and not for new purchases of BID Power Sensors. This option provides for adding the calibration data to a CD and may be ordered in addition to any other required or requested Std Lab Calibration option (must also be ordered in addition to H99). The H99 option is recommended, but not required.