Power Correction Calibration

(M9377A)

This test instructs the operator to load the M9485A VNA soft front panel (SFP) on their PC and perform port configuration, followed by a guided power calibration on the system using a power meter, a power sensor, a cable, and a 50 Ohm load. The corrected power values are then stored.

The port configuration is only required to be performed at the first time the M9485A VNA soft front panel is loaded after the M9377A module is installed, or if there is any change to the configuration.

However, the power calibration MUST be performed prior to any other direct access receiver module performance verification test for results to be considered valid. The calibration is valid for 48 hours after such time it will need to be performed again.

There is also a procedure called User Preset File Settings that is recommended to prevent source unleveled error.

Required Test Equipment

Test Equipment

Recommended Model Number1

Power meter

N1914A

Power sensor

E9304A-H19

Load

Any
(can come from the 85052D cal kit)

  1. Refer to the main equipment list for alternative models.

Procedures and Connection Setups

To expand all, click the Expand All button in the toolbar (webhelp only).

Port Configuration

Only perform this if the Port Configuration Wizard has not been executed before, or if there is any change to the configuration.

The Port Configuration Wizard will be launched automatically when the M9485A VNA soft front panel (SFP) is executed for the first time. If it does, go to Step 3. Otherwise, start from Step 1.

  1. From the M9485A VNA SFP, click System > Service > System Power Cal...

  2. Click Port Config.

  3. Make sure the port configuration is DA, then click Store Port Configuration.

  4. Click Yes.

  5. Click OK and restart the M9485A VNA SFP.

System Power Calibration

  1. From the M9485A VNA SFP, click System > Service > System Power Cal...

  2. Click Calibration.

  3. Click Next.

  4. Select the appropriate power meter VISA address and click Next.

  5. Click Start.

  6. Connect the power sensor to Port 1 Src Out of the DUT and click OK.

    ClosedClick here for images.

     

  7. Connect Port 1 Src Out to Port 1 T In of the DUT with a cable and click OK.

    ClosedClick here for images.

     

  8. Click Next.

  9. Click Start.

  10. Connect the power sensor to Port 2 of the DUT and click OK.

    ClosedClick here for images.

     

  11. Click Next.

  12. Click Start.

  13. Connect a 50 Ohm load to Port 2 of the DUT and click OK.

    ClosedClick here for images.

     

  14. Click Next.

  15. Click Finish.

  16. Click OK and restart the M9485A VNA SFP.

User Preset File Settings

This section is recommended to prevent source unleveled error.

  1. From the M9485A VNA SFP, click Freq > Start. Set the start frequency to 50 MHz.

  2. Click Power > Power Level. Set the power level to 0 dBm.

  3. Click Preset > User Preset....

  4. Select Save current state as User Preset and select the User Preset Enable box. Click OK.