It is not a requirement to open the M9485A soft front panel for this test. |
This test measures the test port match, thru port match, arm port match, insertion loss, coupling factor and directivity of the test coupler of the directional coupler. These parameters are measured using a network analyzer after the network analyzer has performed a full 3-port calibration
|
Test Equipment |
Recommended Model Number1 |
|---|---|
|
Network analyzer |
E5071C Option 4K5 or 4D5 |
|
ECal module |
85093-60010 (85093C-00F) |
|