Df2 Rate (Bluetooth)

This result reports the percentage of measured Δf2max values that exceed the minimum deviation required by the standard. The floor depends on the PHY Physical Layer — for example, 185 kHz kiloHertz: A radio frequency measurement (one kilohertz = one thousand cycles per second). for LE 1M and 370 kHz for LE 2M (RF Radio Frequency: A generic term for radio-based technologies, operating between the Low Frequency range (30k Hz) and the Extra High Frequency range (300 GHz).-PHY Test Specification §4.6.3); the corresponding BR requirement is 115 kHz.

The pass criterion is that at least 99.9% of all Δf2max values, pooled over at least 10 packets, exceed the floor. Read the pooled figure from the Average column with averaging ON.

See Also

Df2 Max

Limits Table

Results Table