Tracking Settings (6GR)

The Tracking section of the Advanced tab controls signal tracking. While 3GPP EVM (Conformance) is selected (the default), these settings are forced to the values the conformance specification requires. While it is selected the section is also collapsed and its header says that conformance fixes these values, and each parameter's tooltip names that gate. Clear 3GPP EVM (Conformance) to re-open the section and control them directly; the defaults below then apply. All parameters are auto-coupled — see Auto and Manual Parameter Coupling.

Tracking Mode - Selects the signal basis used for tracking.

Default: RS+Data (3GPP while conformance is selected)

Choices: RS+Data, RS, 3GPP

Track Amplitude - Enables amplitude tracking.

Default: Selected (forced off while conformance is selected)

Track Phase - Enables phase tracking.

Default: Selected (forced by the conformance specification while it is selected; off for FR1)

Track Timing - Enables timing tracking.

Default: Selected (forced off while conformance is selected)

Compensate Symbol Clock Offset - Compensates for symbol clock offset in the analysis.

Default: Selected (forced off while conformance is selected)

See Also

Advanced Tab

EVM Settings