[:SENSe]:CORRection:COLLect:GUIDed:STEP:PROMpt? <num>

(Read-Only) Returns the cal standard prompt to be used for the specified step number in the guided cal. Use [:SENSe]:CORRection:COLLect:GUIDed:SCOunt to read the number of steps required for the cal.

Use [:SENSe]:CORRection:COLLect:GUIDed:STEP:ACQuire to perform the cal std measurement.

Relevant Modes

CAT, NA

To Cal in VVM mode, perform a Cal in CAT or NA mode, then switch to VVM mode.

Parameters

 

<num>

Step number of the guided calibration.

Examples

CORR:COLL:GUID:STEP:PROM? 1

See Guided Cal Example

Query Syntax

Not Applicable

Default

Not Applicable


Last Modified:

18-Oct-2012

New command