Message 330

Message

"330. The {0} module has corrupt contents. Recovery may be possible."

Description

The indicated plug-in module is exhibiting symptoms of an internal failure. Try removing the module and reinstalling it. With your fingers, apply pressure the top of the module's faceplate while hand tightening the front-panel module locking screws. Cycle power on the XR8 oscilloscope, and contact Keysight if the module still indicates an error.