:CALCulate:COMParator:CONDition{1|2|3|4}:NOMinal <numeric>
:CALCulate:COMParator:CONDition{1|2|3|4}:NOMinal?
Lets you specify the limit range reference value that applies to each of the four sorting conditions (:COND1 through :COND4). The reference value applies to all of the bins if you have issued the :CALC:COMP:COND{1-32}:MODE command by specifying “DEV” or “PCNT” to configure the instrument to accept a limit range in deviation or percent mode.
Parameter |
Value |
Description |
Limit range reference value |
Range |
-1E20 to 1E20 |
Preset value |
0 |
Unit |
Depends on measurement parameter used to sort DUTs. |
Resolution |
1E-20 |
If your specified parameter is beyond the valid range, the instrument ignores the parameter and uses either the maximum or minimum value depending on whether the parameter is greater than the upper limit or smaller than the lower limit.
{numeric}<newline><^END>
Stim Table > Comparator tab > Condition{1-32} > Reference > {numeric}, or
Analysis > Edit Comptr Table > Condition{1-32} > Reference > {numeric}