:CALC:COMP:COND{1-32}:SNUM

Syntax

:CALCulate:COMParator:CONDition{1|2|3|4}:SNUMber <numeric>

:CALCulate:COMParator:CONDition{1|2|3|4}:SNUMber?

Description

Lets you specify the measurement point for each of the four sorting conditions (:COND1 through :COND4). This setting applies to all of the bins.

This setting is reset when settings in the measurement point setup table are changed by execution of :SOUR:LIST or :SOUR:LIST:CLE commands.

Variable

Parameter

Value

Description

Point number

Range

1 to number of points (up to 201)

Preset value

1

Resolution

1

If your specified parameter is beyond the valid range, the instrument ignores the parameter and uses either the maximum or minimum value depending on whether the parameter is greater than the upper limit or smaller than the lower limit.

Query response

{numeric}<newline><^END>

Related objects

:CALC:COMP:COND{1-32}:PAR

:SOUR:LIST

:SOUR:LIST:CLE

Equivalent key

Stim Table > Comparator tab > Stimulus, or

Analysis > Edit Comptr Table > Stimulus