Setting Measurement Parameters

You can assign measurement parameters 1 through 4 (:PAR1, :PAR2, :PAR3, :PAR4) independently of the other parameter settings. To set measurement parameters, use the following command:

You can configure the instrument to measure the parameters shown in the table below:

Parameter

Parameter label on screen

Description

Z

|Z|

Impedance amplitude (absolute value)

Y

|Y|

Admittance amplitude (absolute value)

LS

Ls

Equivalent series inductance

LP

Lp

Equivalent parallel inductance

CS

Cs

Equivalent series capacitance

CP

Cp

Equivalent parallel capacitance

RS

Rs

Equivalent series resistance

RP

Rp

Equivalent parallel resistance

Q

Q

Q value (inverse of dissipation factor)

D

D

Dissipation factor

X

X

Equivalent series reactance

R

R

Resistance

G

G

Equivalent parallel conductance

B

B

Equivalent parallel susceptance

TZR

θz (rad)

Impedance phase (in radians)

TZD

θz (deg)

Impedance phase (in degrees)

TYR

θy (rad)

Admittance phase (in radians)

TYD

θy (deg)

Admittance phase (in degrees)

USER1
to USER4

Γ(db), Γ(deg), I-mon, V-mon (Default)

User parameters

Sample Program

See Measurement Conditions.

Other topics about Setting Measurement Conditions