:SENS<Ch>:CORR2:EDEL:LENG

Type

Command/Query

Syntax

:SENSe<Ch>:CORRection2:EDELay:LENGth <Value>

:SENSe<Ch>:CORRection2:EDELay:LENGth?

Description

This command sets/gets the extension value in meter.

SENS:CORR1:EDEL:LENG causes an undefined header error.

Variable

Parameter

Value

Description

Extension

Data Type

Numeric type (Real)

Range

-

Preset Value

0

Unit

meter

Resolution

-

Query Response

{numeric}<newline><^END>

Examples

Dim Var as Double
Ana.WriteString ":SENS1:CORR2:EDEL:LENG " & Var, True
Ana.WriteString ":SENS1:CORR2:EDEL:LENG?", True
Var=Ana.ReadString

Related Commands

:SENS:CORR:EDEL:STAT

:SENS:CORR:EDEL:TIME

Equivalent Softkey

Cal > Port Extension > Extension (Length)

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION(2).EDELAY.LENGTH