:SENS<Ch>:CORR2:EDEL:TIME

Type

Command/Query

Syntax

:SENSe<Ch>:CORRection2:EDELay:TIME <Value>

:SENSe<Ch>:CORRection2:EDELay:TIME?

Description

This command sets/gets the port extension value in time.

SENS:CORR1:EDEL:TIME causes an undefined header error.

Variable

Parameter

Value

Description

Port extension value

Data Type

Numeric type (Real)

Range

-

Preset Value

0

Unit

sec

Resolution

-

Query Response

{numeric}<newline><^END>

Examples

Dim Var as Double
Ana.WriteString ":SENS1:CORR1:EDEL:TIME ON", True
Ana.WriteString ":SENS1:CORR1:EDEL:TIME?", True
Var=Ana.ReadNumber

Related Commands

:SENS:CORR:EDEL:STAT

:SENS:CORR:EDEL:LENG

Equivalent Softkey

Cal > Port Extension > Extension (Time)

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION(Type).EDELAY.TIME