SCPI.SENSe.UDConverter.STATe

Object Type

Property (Read-Write)

Syntax

SCPI.SENSe.UDConverter.STATe = Value

Value = SCPI.SENSe.UDConverter.STATe

Description

This command sets/gets user downconverter ON/OFF.

Variable

Parameter

   Value

Description

User downconverter (On/Off)

Data Type

Boolean type (Boolean)

Range

ON|OFF|1|0

Preset Value

OFF

Unit

-

Resolution

-

Preset Effect

Yes

Save-Recall Effect

Yes

Error

-

Examples

Dim Var as Boolean
Var= 1
SCPI.SENSe.UDConverter.STATe = Var
Var = SCPI.SENSe.UDConverter.STATe

Related Objects

SCPI.SENSe.UDConverter.HARMonic

SCPI.SENSe.UDConverter.LO

SCPI.SENSe.UDConverter.MODE

Equivalent Key

PN Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

SP Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

FP Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

TR Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

AM Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

BB Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

USER Menu: System > Instrument Setup > Frequency Offset (User Downconv.) > Frequency Offset

Equivalent SCPI Command

Syntax

:SENSe:UDConverter:STATe {ON|OFF|1|0}

:SENSe:UDConverter:STATe?

Query Response

{ON|OFF|1|0} <newline><^END>

Example of use

10 OUTPUT 717;":SENSe:UDConverter:STATe 1"
20 OUTPUT 717;":SENSe:UDConverter:STATe ?"
30 ENTER 717;A