Type topic text here.
TRIGger:CHANnel :TR | VIDeo | NARRow[1-2] | FREQuency | HOFF | POWer | WIDE | FREQuency | HOFF
|
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the adjustment time from the external trigger. Command only. This feature is not available on the Soft Front Panel (GUI). |
|
Parameters |
|
<value> |
Adjustment time (from 0 to 1E-6 sec.) |
Examples |
TRIG:CHAN:TR:ETTA 1E-6 trigger:channel2:tr:ettadjust 1E-6 |
Query Syntax |
TRIGger:CHANnel<cnum>:TR:ETTAdjust? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with opt S963015B (Read-Write) Enable and disable the video trigger for the Transient measurement. |
|
Parameters |
|
<Boolean> |
Choose from: OFF / 0 ON / 1 |
Examples |
TRIG:CHAN:TR:VID:ENAB ON trigger:channel:tr:video:enable off |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:ENABle? |
Return Type |
Boolean |
OFF |
Applicable Models: E505x with opt S963015B (Read-Write) Set and read the hysteresis value (frequency) for video trigger in Narrow band. |
|
Parameters |
|
<value> |
Hysteresis value (frequency) in Hz |
Examples |
TRIG:CHAN:TR:VID:NARR2:FREQ:HYST 2E3 trigger:channel:tr:video:narrow1:frequency:hysteresis 3E3 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:FREQuency:HYSTeresis? |
Return Type |
Numeric |
1E3 |
Applicable Models: E505x with opt S963015B (Read-Write) Set and read the target frequency value which make a video trigger in Narrow band |
|
Parameters |
|
<value> |
Target frequency value in Hz |
Examples |
TRIG:CHAN:TR:VID:NARR1:FREQ:TARG 1E9 trigger:channel:tr:video:narrow2:frequency:target 1E9 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:FREQuency:TARGet? |
Return Type |
Numeric |
1E6 |
Applicable Models: E505xAwith opt S963015B (Read-Write) Set and read the holdoff time for Video Trigger in Narrow band. |
|
Parameters |
|
<value> |
holdoff time in second |
Examples |
TRIG:CHAN:TR:VID:NARR[1-2]:HOFF[:TIME] trigger:channel:tr:video:narrow[1-2]:hoff[:time] |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:HOFF[:TIME]? |
Return Type |
Numeric |
1 |
Applicable Models: E505xAwith opt S963015B (Read-Write) Set and read the hysteresis value (Power) for video trigger in Narrow band. |
|
Parameters |
|
<value> |
Hysteresis value in dBm |
Examples |
TRIG:CHAN:TR:VID:NARR1:POW:HYST trigger:channel:tr:video:narrow2:power:hysteresis |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:POWer:HYSTeresis? |
Return Type |
Numeric |
10 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the target power value which make a video trigger in Narrow band. |
|
Parameters |
|
<value> |
Target power value in dBm |
Examples |
TRIG:CHAN:TR:VID:NARR1:POW:TARG -10 trigger:channel:tr:video:narrow2:power:target -5 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:POWer:TARGet? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the trigger edge from positive or negative for video trigger in Narrow band. |
|
Parameters |
|
<type> |
Choose from: POSitive NEGative |
Examples |
TRIG:CHAN:TR:VID:NARR1:SLOP POS trigger:channel:tr:video:narrow2:slope negative |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:SLOPe? |
Return Type |
Character |
POS |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the trigger type for Narrow1 or Narrow2. |
|
Parameters |
|
<type> |
Choose from: FREQuency - video trigger type is frequency POWer - video trigger type is power |
Examples |
TRIG:CHAN:TR:VID:NARR[1-2]:TYPE FREQ trigger:channel:tr:video:narrow[1-2]:type power |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:TYPE? |
Return Type |
Character |
POWer |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the minimum power value which disables the video trigger in Narrow band. The trigger is disabled if the power level is below this level. The relative value from the maximum input level should be specified. |
|
Parameters |
|
<value> |
Minimum Power value in dB |
Examples |
TRIG:CHAN:TR:VID:NARR1:THR: -5 trigger:channel:tr:video:narrow2:threshold:value -10 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:NARRow[1-2]:THReshold[:VALue]? |
Return Type |
Numeric |
-20 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the trigger source from Wide, Narrow1 or Narrow2. |
|
Parameters |
|
<type> |
Choose from: WB - video trigger source is WB NB1 - video trigger source is NB1 NB2 - video trigger source is NB2 |
Examples |
TRIG:CHAN:TR:VID:SOUR NB1 trigger:channel:tr:video:source nb2 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:SOURce? |
Return Type |
Character |
WB |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the hysteresis value for video trigger in Wideband. |
|
Parameters |
|
<value> |
hysteresis value in Hz |
Examples |
TRIG:CHAN:TR:VID:WIDE:FREQ:HYST 1E6 trigger:channel:tr:video:wide:frequency:hysteresis 1E6 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:WIDE:FREQuency:HYSTeresis? |
Return Type |
Numeric |
1E3 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the target frequency value which make a video trigger in Wideband. |
|
Parameters |
|
<value> |
target frequency value in Hz |
Examples |
TRIG:CHAN:TR:VID:WIDE:FREQ:TARG 2E6 trigger:channel:tr:video:wide:frequency:target 2E6 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:WIDE:FREQuency:TARGet? |
Return Type |
Numeric |
1E6 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the holdoff time for Video Trigger in Wide band. |
|
Parameters |
|
<value> |
Holdoff time in second |
Examples |
TRIG:CHAN:TR:VID:WIDE:HOFF 0.1 trigger:channel:tr:video:wide:hoff:time 0.3 |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:WIDE:HOFF[:TIME]? |
Return Type |
Numeric |
1 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the trigger edge from positive or negative for video trigger in Wideband. |
|
Parameters |
|
<type> |
Choose from: POSitive NEGative |
Examples |
TRIG:CHAN:TR:VID:WIDE:SLOP POS trigger:channel:tr:video:wide:slope negative |
Query Syntax |
TRIGger:CHANnel:TR:VIDeo:WIDE:SLOPe? |
Return Type |
Character |
POSitive |