:SENS<Ch>:CORR:COLL:SAVE

Type

Command

Syntax

:SENSe<Ch>:CORRection:COLLect:SAVE

Description

This command calculates the calibration coefficients depending on the calibration type selection, from the measured calibration data.

Calculating the calibration coefficients clears all the measured calibration data whether or not used for the calculation and also clears the calibration type selection.

If this command is executed before all necessary calibration data for calculating the calibration coefficients is measured, an error occurs and the command is ignored.

Examples

Ena.WriteString ":SENS:CORR:COLL:SAVE", True

 

See Calibration.

Related Commands

:SENS:CORR:COLL:ISOL

:SENS:CORR:COLL:LOAD

:SENS:CORR:COLL:OPEN

:SENS:CORR:COLL:SHOR

:SENS:CORR:COLL:THRU

:SENS:CORR:COLL:TRLL

:SENS:CORR:COLL:TRLR

:SENS:CORR:COLL:TRLT

Equivalent Softkey

Cal > Calibrate > {Response |n-Port Cal } > Done

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION.COLLECT.SAVE