Perform a 4-Port Cal with One 2-Port ECal Module


You can perform a 4-port calibration with a 2-Port ECal Module. When all four DUT connectors are the same type and gender, the calibration can occur with only four connections, the same number of connections you would make with a 4-port ECal module.

Important Note: DO NOT connect/disconnect USB devices during ECal calibrations. Doing so may cause problems with the calibration.

Procedure

  1. Connect the 2-port ECal module to a VNA USB port.

  2. Press Cal > Main > Other Cals > Smart Cal....

  3. Select 4 Port Cal, then click Next to see the following dialog:

  4. Select the DUT Connectors for each port. In this example, all four DUT connectors are Type N, female.

  5. Select the attached ECal module. We are using a 85092-60007 ECal module.

  6. Select Modify Cal (Show Advanced Settings for ECal) then click Next to see the following dialog:

  7. For the fewest number of physical connections, select the default port assignments.

  8. Select ECal Thru as Unknown. This is the most accurate and easiest Thru Cal Method. Learn more.

  9. You may need to clear Do Orientation when calibrating at low power levels. Learn more. This will add additional connection steps.

  10. Follow the prompts to complete the calibration:

    1. Connect ECal to ports 1 and 2. Click Measure.

    2. Connect ECal to ports 1 and 3. Click Measure.

    3. Connect ECal to ports 1 and 4. Click Measure.

  11. At the Specify Delay dialogs, click OK. This is the measured delay for each of the Thru connections in the ECal module. Learn more.

  12. Click either Save As User Cal Set, or Finish.