Real Time

1. System Requirements

Installation Steps: 1 2 3 4 5 6 7

To generate signals using the N7601B Signal Studio for cdma2000/1xEV-DO software, you must have a properly configured PC and a properly configured instrument, as described in the following table.

Instrument Requirements

N5172B EXG X-Series RF vector signal generator

Firmware:

Version B.01.10 or later (how to upgrade

Required options:

N5172B-503 (9 kHz to 3 GHz frequency range)

N5172B-653 (Internal baseband generator, 32 MSa/sec, up to 60 MHz RF bandwidth)

N5172B-660 (upgrade baseband generator with real-time capability)

Recommended options:

N5172B-009 (internal Flash drive)

N5172B-021 (upgrade baseband generator memory from 32 MSa to 256 MSa)

N5172B-1EA (high power output for PA testing)

N5172B-1ER (flexible reference input, 1-50 MHz)

N5172B-403 (calibrated AWGN for testing with noise)

N5172B-655 (upgrade baseband generator from 60 to 120 MHz RF bandwidth)

N5172B-UNV (enhanced dynamic range)

N5172B-UNZ (fast switching for manufacturing environments)

N5182B MXG X-Series RF vector signal generator

Firmware:

Version B.01.10 or later (how to upgrade)

Required options:

N5182B-503 (9 kHz to 3 GHz frequency range)

N5182B-656 (Internal baseband generator, 32 MSa/sec, up to 80 MHz RF bandwidth)

N5182B-660 (upgrade baseband generator with real-time capability)

Recommended options:

N5182B-009 (internal Flash drive)

N5182B-022 (upgrade baseband generator memory from 32 MSa to 512 MSa)

N5182B-1EA (high power output for PA testing)

N5182B-1ER (flexible reference input, 1-50 MHz)

N5182B-1EQ (low power output for Rx testing)

N5182B-403 (calibrated AWGN for testing with noise)

N5182B-657 (upgrade baseband generator from 80 to 160 MHz RF bandwidth)

N5182B-UNV (enhanced dynamic range)

N5182B-UNZ (fast switching for manufacturing environments)

M9381A PXIe vector signal generator

Driver:

Version 1.4.400.0 or later

Recommended options:

M9381A-M05 (memory 512 MSa)

M9381A-M10 (memory 1024 MSa)

M9381A-1EA (high output power)

M9381A-B10 (RF modulation bandwidth 100 MHz)

M9381A-B16 (RF modulation bandwidth 160 MHz)

Additional options to consider:

M9381A-F06 (frequency range 1 MHz to 6 GHz)

Limitations:

The M9381A VSG has a limitation related to the Power Search and DC Cal functions, creating a time-synchronization loss. That is, the signal timing changes after a Power Search or DC Cal. When ALC is set to Off, Signal Studio automatically runs Power Search, which affects the user's trigger signal for starting waveform playback. Even though the timing alignment of the trigger signal and waveform are properly set, the Power Search function disrupts this synchronization. In these cases, use the following workarounds:

  • Set ALC to On whenever possible. This avoids the automatic Power Search.
  • When ALC must be set to Off (or is set to Off by Signal Studio), you can re-trigger the signal by turning ARB Off, then On, and then send the trigger signal to M9381A VSG.

For DC Cal, run it before you download and play the waveform. Do not run DC Cal while the signal is playing, which disrupts the synchronization.

E6640A EXM wireless test set

Software:

Version E.16.06 or later

Recommended options:

E6640A-B85 (increases signal bandwidth to 80 MHz)

E6640A-B1X (increases signal bandwidth to 160 MHz)

Y-Options (these option models, beginning with "Y," combine signal creation and measurement applications)

Additional options to consider:

E6640A-506 (380 MHz to 6 GHz frequency range)

E6640A-5WC (frequency range for wireless connectivity)

N5182A MXG Vector Signal Generator - requires firmware version A.01.10 or later (how to upgrade)

N5182A-503

250 kHz to 3 GHz frequency range

Or the other frequency option:

     N5182A-506

250 kHz to 6 GHz frequency range

N5182A-651

Internal Baseband Generator (30 MSa/s and 8 MSa)

Or one of the Closedother internal baseband generator options:

Simple guidelines for deciding which internal baseband generator you will need is using the following equation:

Sample Size = ARB Sample Clock X Oversampling Ratio X Number of Carriers

As example,

  • If you have an ARB Sample Clock rate of 1.2288 MHz and you are testing with 4 carriers and an oversampling ratio of 4,

    Sample Size

    = 1.2288 MHz X 4 X 4

    = 19.6608 MSa/s

    Since 19.6608 MSa/s is less than the 30 MSa/s capability of option 651, this internal baseband generator should be adequate for this testing.

  • If you have an ARB Sample Clock rate of 1.2288 MHz and you are testing with 10 carriers and an oversampling ratio of 4,

    Sample Size

    = 1.2288 MHz X 4 X 10

    = 49.152 MSa/s

    Since 49.152 MSa/s is greater than the 30 MSa/s capability of option 651, this internal baseband generator will not be adequate for this testing. However, option 652 has 60 MSa/s capability. So the option 652 internal baseband generator should be adequate for this testing.

     N5182A-652

Internal Baseband Generator (60 MSa/s and 8 MSa)

     N5182A-654

Internal Baseband Generator (125 MSa/s and 8 MSa)

N5182A-UNV

Enhanced Dynamic Range

N5182A-019

Increased baseband generator memory to 64 MSa is also available, in addition to one of the Internal Baseband Generator options listed above. (Used to save several types of waveform files)

N5182A-1EL

Differential I/Q Outputs (Depends on the test requirements)

N5182A-403

Calibrated AWGN Personality (used only for receiver test)

E4438C Vector Signal Generator - requires firmware version C.04.84 or later (how to upgrade)

E4438C-503

250 kHz to 3 GHz frequency range

Or the other frequency option:

     E4438C-506

250 kHz to 6 GHz frequency range

E4438C-6021

Internal Baseband Generator (64 MSa memory)

E4438C-005

6 GB internal hard drive

E4438C-UNJ

Enhanced phase noise (required with E4438C-506)

E4438C-403

Calibrated AWGN Personality (used only for receiver test)

E8267D Vector Signal Generator - requires firmware version C.04.85 or later (how to upgrade)

E8267D-520

250 kHz to 20 GHz frequency range

E8267D-6021

Internal Baseband Generator (64 MSa memory)

E8267D-005

6 GB internal hard drive

E8267D-UNX

Ultra phase noise

E8267D-005

6 GB internal hard drive

E8267D-1EH

Improved harmonics below 2 GHz

E8267D-403

Calibrated AWGN Personality (used only for receiver test)

  N5106A PXB Baseband Generator and Channel Emulator

         N5106A-612

2 DSP blocks on 1 baseband card

         N5106A-632

2 I/O channels - 2 analog I/Q out and 2 digital I/O

         N5106A-EFP

Baseband Generation

         N5106A-JFP

Calibrated AWGN

1. Recommended option: the baseband generator option can be any of the following: E4438C-001, -002, -601, -602 or E8267D-602

PC Requirements

200 GB HDD with 2 GB free disk space at runtime (5 GB or higher recommended)

1024 x 768 screen resolution (1280 x 1024 recommended)

Windows 8.1 (Pro/Enterprise) (32/64 bit)1,2:

Windows 7 (Professional/Ultimate/Enterprise)1,2:

Installed GPIB IO interface card, LAN interface card, or USB/GPIB interface connector

Keysight VISA

Microsoft Windows Installer version 2.0 or later

Adobe Flash Player 7 or later (required for using the Signal Studio Configuration Assistant)


1 No support for Itanium64.

2 Signal Studio is compatible with the 64-bit operating system, but performs at the 32-bit speed.