2. Configuring Parameters in the N7614C Software

The following figure shows the block view settings in N7614C for a dual-band ET measurement.

The remaining section of this topic provides an example procedure of setting N7614C for DPD measurement. Note that for different PA, the settings are different. But there is normally total 10 typical steps to set all related parameter before executing measurement.

Step 1: Set the Waveform Blocks

Step 2: Set the Crest Factor Reduction Blocks

Step 3: Set the Digital Predistortion Block

Step 4: Set the RF Signal Generator Blocks

Step 5: Set the PA Block

Step 6: Set the Signal Analyzer Block

Step 7: Set the Measurement View

Step 8: Set the Envelope Signal Generator Block

Step 9: Set the ETPS Block

Step 10: Set the Envelope Tracking Block

Step 1: Set the Waveform Blocks

  1. Click on the Waveform 1 block icon on the UI. In the Closedright panel, select Preloaded as the Waveform type and select the "LTE_FDD_5M_QPSK_FullRB.wfm" file, set the Bandwidth to 5 MHz according to the waveform bandwidth.

  2. Click on the Waveform 2 block icon on the UI. In the Closedright panel, select Preloaded as the Waveform Type and select the "LTE_FDD_5M_QPSK_PartialRB.wfm" file, set the Bandwidth to 5 MHz according to the waveform bandwidth.

If you want to use a waveform generated by another Keysight signal studio, set Waveform Type to Signal Studio and select the file.

Note that to use a waveform generated by other signal studio software for DPD test, the instrument needs to have the corresponding license. For this example, to use LTE waveform, the N7624B license needs to be installed in the instrument

To perform dual-band DPD measurement, the two waveforms selected for the two bands should have equal time duration; otherwise, DPD measurement will abort.

Step 2: Set the Crest Factor Reduction Blocks

In both Crest Factor Reduction blocks, set the switch to Thr (through).

Step 3: Set the Digital Predistortion Block

  1. In the Digital Predistortion block set the switch to On.

  2. In the Closedright panel, do the following:

Different power amplifiers may have different characteristics, which will need different DPD models to simulate the DPD performance. In DPD test, if the measurement results are not as expected, you can try to change the parameters for DPD model extraction.

Step 4: Set the RF Signal Generator Block

Click the RF Signal Generator block. In the right panel, do the following:

Step 5: Set the PA Block

Click the PA block. In the right panel, do the following:

Step 6: Set the Signal Analyzer Block

Click the Signal Analyzer block. In the right panel, do the following:

Step 7: Set the Measurement View

Click the button to change the current view from Closedblock view to Closedmeasurement view.

Step 8: Set the Envelope Signal Generator Block

Click the Envelope Signal Generator block. In the Closedright panel, do the following:

Refer to Voltage Mapping in Envelope Tracking Measurement for more details about Voltage Mapping.

Step 9: Set the ETPS Block

Click the ETPS block. In the Closedright panel, do the following:

  1. Set the ETPS Gain to 7.0 dB.

  2. Set the Vcc Offset to 2.75 V.

  3. Set the Vcm to 450 mV.

Refer to Voltage Mapping in Envelope Tracking Measurement for more details about Voltage Mapping.

Step 10: Set the Envelope Tracking Block

Switch the Envelope Tracking block to On. In the Closedright panel, do the following: