FETCh:TFDTune

Last updated: 2010-03-10

For more information on the GPIB Command conventions, see GPIB Command conventions

FETCh:TFDTune[:ALL]?

Function

Queries the fast device tune measurement results.

Results are returned in the following order:

  • Integrity Indicator

  • Number of Tx Frequency Steps

  • Number of Tx Power Steps

  • Number of MS TX power steps measured (in most cases, it equals Number of Tx Frequency Steps multiplied by Number of Tx Power Steps)

  • Comma-separated list of 400 MS Tx Power results. 400 results are always returned; if the test set measured fewer than 400 power steps from the mobile, the remaining values are NANs (9.91E+37). The power results for the first frequency in the MS TX Frequency Sequence are returned first, then the power results for the second frequency in the MS TX Frequency Sequence, etc., followed by NANs if fewer than 400 points were measured.

Query

Integrity

Number of Tx Frequency Steps

  • Range: 1 to 40, 9.91 E+37

  • Resolution: 1

Number of Tx Power Steps

  • Range: 1 to 40, 9.91 E+37

  • Resolution: 1

Number of MS Tx Power Steps Measured

  • Range: 1 to 400, 9.91 E+37

  • Resolution: 1

MS Tx Power Results Array

  • Range: -100.00 to +100.00 dBm, 9.91 E+37

  • Resolution: 0.01 dBm

Requirements

Test Application Revision: A.04.00 and above

Programming Example

"FETCh:TFDT?"

FETCh:TFDTune:INTegrity?

Function

Queries the measurement integrity indicator for the last FDT measurement completed.

Zero indicates a normal measurement. See Integrity Indicator for information about non-zero integrity results.

Query

Range: See Integrity Indicator

Requirements

Test Application Revision: A.04.00 and above

Programming Example

"FETCh:TFDTune:INTegrity?"