ABORt

ABORt

Last updated: January 5, 2010

GPIB Command Conventions

ABORt[:ALL]

Function

Stops all measurements that are active.

If the trigger arm is set to single, the measurements will go to the idle state.

If the trigger arm is set to continuous, the measurements will re-arm and initiate again.

Requirements

Test or Lab Application Revision: A.01 and above

Programming Example

 
OUTPUT 714;"ABORt"
! Aborts all active measurements in progress.

ABORt:<measurement mnemonic>

Function

Stops the selected measurement if it is active.

If the trigger arm is set to single, the measurements will go to the idle state.

If the trigger arm is set to continuous the measurements will re-arm and initiate again.

  • AFANalyzer - Audio Frequency Analyzer measurements
  • CFDTune - Fast Device Tune measurement ( lab application or feature-licensed test application only )
  • EVMQpsk - QPSK EVM measurement ( lab application only )
  • FSTability - Frequency Stability measurement
  • HBLerror - HSDPA Block Error Ratio ( lab application or feature-licensed test application only )
  • HRCQuality - HSDPA Reporting of Channel Quality Indicator (lab application only)
  • SAUDio - Swept Audio measurement
  • SMONitor - Spectrum Monitor measurement
  • WACLeakage - Adjacent Channel Leakage
  • WBERror - Loopback Bit Error Ratio measurement
  • WBLerror - Loopback Block Error Ratio measurement
  • WCDomain - Code Domain measurements
  • WCPower - Channel Power measurement
  • WCTFormat - Change of TFC ( lab application only )
  • WDPAnalysis - Dynamic Power Analysis
  • WEDPower - Extended Range Dynamic Power
  • WILPower - Inner Loop Power measurement
  • WIQTuning - IQ Tuning measurement
  • WOBWidth - Occupied Bandwidth measurement
  • WOOPower - PRACH Transmit On/Off Power measurement
  • WOOSynch - Out-of-Synch Handling of Output Power
  • WPDiscon - Phase Discontinuity
  • WPESquality - Perceptual Evaluation of Speech Quality measurement
  • WPPAnalysis - PRACH Preamble Analysis ( lab application only )
  • WSEMask - Spectrum Emission Mask measurement
  • WTDPower - TX Dynamic Power measurement
  • WTPower - Thermal Power measurement
  • WWQuality - Waveform Quality measurement

Requirements

Test Application Revision A.01 and above.

Test Application Revision A.02 added WCDomain, WSEMask, and WOBWidth.

Test Application Revision A.03 added WIQTuning, WILPower, WOOPower, and SMONitor.

Test Application Revision A.05 added WBLerror, WDPAnalysis and WPDiscon.

Test Application Revision A.06 added FSTability

Test Application Revision A.07 added WTDPower

Test Application Revision A.08 with the required feature license, added HBLerror

Test Application Revision A.09 with the required feature license, added CFDTune

Lab Application Revision: A.01 and above

Lab Application Revision B.01 added WBLerror, WDPAnalysis, WPDiscon, and WPPAnalysis

Lab Application Revision C.01 added FSTability, WCTFormat and WOOSynch

Lab Application Revision C.02 added WTDPower

Lab Application Revision C.03 added HBLerror

Lab Application Revision C.04 and D.00.04 added CFDTune

Test Application Revision A.10 and Lab Application Revision D.01 and above, added WEDPower

Lab Application Revision E.00.16 and above added HRCQuality

Lab Application Revision F.00 added EVMQpsk

Lab Application Revision G.00 added WPESquality

Programming Example

 
OUTPUT 714;"ABORt:WWQuality"
! Aborts the waveform quality measurements.