CALL:SSCell

CALL:SSCell

Last updated: April 27, 2013

This topic is only applicable to the lab application.

GPIB Command Conventions

CALL:SSCell:CHANnel:ADJacent

Function

Sets/queries the downlink adjacent UARFCN of the secondary serving cell.

In 3GPP Rel'8, the UARFCN of the secondary serving cell must be adjacent to the serving cell.

The DL UARFCN of the secondary serving cell must be in same frequency range with the currently active band.

This setting is only allowed to be changed when the Call Status is Idle .

Setting

Range: HIGHer | LOWer

Query

Range: HIGH | LOW

*RST Setting

HIGHer

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CHANnel:ADJacent HIGHer"

CALL:SSCell:CHANnel:CONFig[:TYPE]

Function

Sets/queries Secondary Serving Cell DL Channel Config .

This setting is only allowed to be changed when the Call/Data Status is Idle .

Setting

Range: ADJacent| MANual

Query

Range: ADJacent| MANual

*RST Setting

ADJacent

Requirements

Test Application Revision: A.20.00 and above, with the required feature license

Lab Application Revision: I.00.00 and above, with the required feature license

Programming Example

 
OUTPUT 714;"CALL:SSCell:CHANnel:CONFig:TYPE MANual"

CALL:SSCell:CHANnel:MANual:UARFcn

Function

Sets/queries the Secondary Serving Cell DL M anual UARFCN .

Setting

Range:

  • 412, 437, 462, 487, 512, 537, 562, 587, 612, 637, 662, 687
  • 712 to 763
  • 787, 812, 837
  • 862 to 912
  • 1007, 1012, 1032, 1037, 1062, 1087
  • 1162 - 1513, resolution = 1
  • 1537 - 1738, resolution = 1
  • 1887, 1912, 1937, 1962, 1987, 2012, 2037, 2062, 2087
  • 2237 - 2563, resolution = 1
  • 2587, 2612, 2637, 2662, 2687, 2712, 2737, 2762, 2787, 2812, 2837, 2862, 2887, 2912
  • 2937 - 3088, resolution = 1
  • 3112 to 3388, resolution = 1
  • 3412, 3437, 3462, 3487, 3512, 3537, 3562, 3587, 3612, 3637, 3662, 3687
  • 3837 - 3903, resolution = 1
  • 3927, 3932, 3957, 3962, 3987, 3992
  • 4017 - 4043, resolution = 1
  • 4067, 4092
  • 4117 - 4143, resolution = 1
  • 4167, 4192
  • 4357 - 4458, resolution = 1
  • 4512 to 4638
  • 9237 - 9387, resolution = 1
  • 9662 - 9938, resolution = 1
  • 10562 - 10838, resolution = 1

Query

Same as setting

*RST Setting

3012
Requirements

Test Application Revision: A.20.00 and above, with the required feature license

Lab Application Revision: I.00.00 and above.

Programming Example

 
OUTPUT 714;"CALL:SSCell:CHANnel:MANual:UARFcn 3012"

CALL:SSCell:CHANnel:MANual:BARBitrator

Function

Sets/queries the Secondary Serving Cell M anual Band Arbitrator .

Setting

Range: BAND5|BAND6

Query

Same as setting

*RST Setting

BAND6

Requirements

Test Application Revision: A.20.00 and above, with the required feature license

Lab test Application Revision: I.00.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CHANnel:MANual:BARBitrator BAND5"

CALL:SSCell:CONNected:CPIChannel:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected CPICH Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:LEVel:HSDPa -10"

CALL:SSCell:CONNected:CPIChannel:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected CPICH State.

Setting

Range: 1|ON|0|OFF

Query

Same as setting

*RST Setting

1|ON

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:STATe:HSDPa On"

CALL:SSCell:CONNected:CPIChannel[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected CPICH Power Offset (see CALL:SSCell:CONNected:CPIChannel:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected CPICH State to on (see CALL:SSCell:CONNected:CPIChannel:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:HSDPa -10"

CALL:SSCell:CONNected:CPIChannel:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected CPICH Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:LEVel:HSPA -10"

CALL:SSCell:CONNected:CPIChannel[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected CPICH Power Offset (see CALL:SSCell:CONNected:CPIChannel:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected CPICH State to on (see CALL:SSCell:CONNected:CPIChannel[:SLEVel]:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:HSPA -10"

CALL:SSCell:CONNected:CPIChannel:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected CPICH State.

Setting

Range: 1|ON|0|OFF

Query

Same as setting

*RST Setting

1|ON

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CPIChannel:STATe:HSPA On"

CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected P-CCPCH/SCH Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-12

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSDPa -12"

CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected P-CCPCH/SCH State

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSDPa On"

CALL:SSCell:CONNected:CCPChannel:PRIMary[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected P-CCPCH/SCH Power Offset (see CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected P-CCPCH/SCH State to on (see CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-12

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:HSDPa -12"

CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected P-CCPCH/SCH Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-12

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSPA -12"

CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected P-CCPCH/SCH State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSPA On"

CALL:SSCell:CONNected:CCPChannel:PRIMary[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected P-CCPCH/SCH Power Offset (see CALL:SSCell:CONNected:CCPChannel:PRIMary:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected P-CCPCH/SCH State to on (see CALL:SSCell:CONNected:CCPChannel:PRIMary:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-12

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:CCPChannel:PRIMary:HSPA -12"

CALL:SSCell:CONNected:PICHannel:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected PICH Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:LEVel:HSDPa -15"

CALL:SSCell:CONNected:PICHannel:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected PICH State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:STATe:HSDPa On"

CALL:SSCell:CONNected:PICHannel[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected PICH Power Offset (see CALL:SSCell:CONNected:PICHannel:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected PICH State to on (see CALL:SSCell:CONNected:PICHannel:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:HSDPa -15"

CALL:SSCell:CONNected:PICHannel:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected PICH Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:LEVel:HSPA -15"

CALL:SSCell:CONNected:PICHannel:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected PICH State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:STATe:HSPA On"

CALL:SSCell:CONNected:PICHannel[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected PICH Power Offset (see CALL:SSCell:CONNected:PICHannel:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected PICH State to on (see CALL:SSCell:CONNected:PICHannel:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:PICHannel:HSPA -15"

CALL:SSCell:CONNected:HSPDschannel:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-PDSCH Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-3

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:LEVel:HSDPa -3"

CALL:SSCell:CONNected:HSPDschannel:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-PDSCH State

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:STATe:HSDPa On"

CALL:SSCell:CONNected:HSPDschannel[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-PDSCH Power Offset (see CALL:SSCell:CONNected:HSPDschannel:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-PDSCH State to on (see CALL:SSCell:CONNected:HSPDschannel:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-3

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:HSDPa -3"

CALL:SSCell:CONNected:HSPDschannel:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-PDSCH Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-3

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:LEVel:HSPA -3"

CALL:SSCell:CONNected:HSPDschannel:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-PDSCH State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:STATe:HSPA On"

CALL:SSCell:CONNected:HSPDschannel[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-PDSCH Power Offset (see CALL:SSCell:CONNected:HSPDschannel:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-PDSCH State to on (see CALL:SSCell:CONNected:HSPDschannel:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-3

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSPDschannel:HSPA -3"

CALL:SSCell:CONNected:HSSCchannel[1]:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 1 Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel[1]:LEVel:HSDPa -10"

CALL:SSCell:CONNected:HSSCchannel[1]:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 1 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel:STATe:HSDPa On"

CALL:SSCell:CONNected:HSSCchannel[1][:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 1 Power Offset (see CALL:SSCell:CONNected:HSSCchannel[1]:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 1 State to on (see CALL:SSCell:CONNected:HSSCchannel[1]:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel:HSDPa -10"

CALL:SSCell:CONNected:HSSCchannel[1]:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 1 Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel:LEVel:HSPA -10"

CALL:SSCell:CONNected:HSSCchannel[1]:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 1 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel:STATe:HSPA On"

CALL:SSCell:CONNected:HSSCchannel[1][:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 1 Power Offset (see CALL:SSCell:CONNected:HSSCchannel[1]:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 1 State to on (see CALL:SSCell:CONNected:HSSCchannel[1]:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel:HSPA -10"

CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 2 Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSDPa -15"

CALL:SSCell:CONNected:HSSCchannel2:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 2 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:STATe:HSDPa Off"

CALL:SSCell:CONNected:HSSCchannel2[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 2 Power Offset (see CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 2 State to on (see CALL:SSCell:CONNected:HSSCchannel2:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:HSDPa -10"

CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 2 Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSPA -15"

CALL:SSCell:CONNected:HSSCchannel2:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 2 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:STATe:HSPA Off"

CALL:SSCell:CONNected:HSSCchannel2[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 2 Power Offset (see CALL:SSCell:CONNected:HSSCchannel2:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 2 State to on (see CALL:SSCell:CONNected:HSSCchannel2:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel2:HSPA -10"

CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 3 Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSDPa -15"

CALL:SSCell:CONNected:HSSCchannel3:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 3 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:STATe:HSDPa Off"

CALL:SSCell:CONNected:HSSCchannel3[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 3 Power Offset (see CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 3 State to on (see CALL:SSCell:CONNected:HSSCchannel3:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:HSDPa -10"

CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 3 Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSPA -15"

CALL:SSCell:CONNected:HSSCchannel3:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 3 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:STATe:HSPA Off"

CALL:SSCell:CONNected:HSSCchannel3[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 3 Power Offset (see CALL:SSCell:CONNected:HSSCchannel3:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 3 State to On (see CALL:SSCell:CONNected:HSSCchannel3:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel3:HSPA -15"

CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 4 Power Offset.

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSDPa -15"

CALL:SSCell:CONNected:HSSCchannel4:STATe:HSDPa

Function

Sets/queries the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 4 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:STATe:HSDPa Off"

CALL:SSCell:CONNected:HSSCchannel4[:SLEVel]:HSDPa

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 4 Power Offset (see CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSDPa ) and (2) sets the UTRA FDD (HSDPA) Secondary Serving Cell Connected HS-SCCH 4 State to on (see CALL:SSCell:CONNected:HSSCchannel4:STATe:HSDPa ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:HSDPa -15"

CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 4 Power Offset

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSPA -15"

CALL:SSCell:CONNected:HSSCchannel4:STATe:HSPA

Function

Sets/queries the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 4 State.

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:STATe:HSPA Off"

CALL:SSCell:CONNected:HSSCchannel4[:SLEVel]:HSPA

Function

The setting form of this command performs two functions: (1) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 4 Power Offset (see CALL:SSCell:CONNected:HSSCchannel4:LEVel:HSPA ) and (2) sets the UTRA FDD (HSPA) Secondary Serving Cell Connected HS-SCCH 4 State to on (see CALL:SSCell:CONNected:HSSCchannel4:STATe:HSPA ).

Setting

Range: -20.00 to 0 dB

Resolution 0.01

Query

Same as setting

*RST Setting

-15

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:CONNected:HSSCchannel4:HSPA -15"

CALL:SSCell:HSSCchannel1:CCODe:CODE

Function

Sets/queries the Secondary Serving Cell HS-SCCH 1 Channelization Code.

Code 0 is not allowed as it would collide with the CPICH and SCH/PCCPCH channels.

This setting cannot be changed while the Operating Mode is Active Cell .

Setting

Range: 1-127

Query

Same as setting

*RST Setting

2

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSCchannel1:CCODe:CODE 2"

CALL:SSCell:HSSCchannel2:CCODe:CODE

Function

Sets/queries the Secondary Serving Cell HS-SCCH 2 Channelization Code.

Code 0 is not allowed as it would collide with the CPICH and SCH/PCCPCH channels.

This setting cannot be changed while the Operating Mode is Active Cell .

Setting

Range: 1-127

Query

Same as setting

*RST Setting

6

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSCchannel2:CCODe:CODE 6"

CALL:SSCell:HSSCchannel3:CCODe:CODE

Function

Sets/queries the Secondary Serving Cell HS-SCCH 3 Channelization Code.

Code 0 is not allowed as it would collide with the CPICH and SCH/PCCPCH channels.

This setting cannot be changed while the Operating Mode is Active Cell .

Setting

Range: 1-127

Query

Same as setting

*RST Setting

9

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSCchannel3:CCODe:CODE 9"

CALL:SSCell:HSSCchannel4:CCODe:CODE

Function

Sets/queries the Secondary Serving Cell HS-SCCH 4 Channelization Code.

Code 0 is not allowed as it would collide with the CPICH and SCH/PCCPCH channels.

This setting cannot be changed while the Operating Mode is Active Cell .

Setting

Range: 1-127

Query

Same as setting

*RST Setting

10

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSCchannel3:CCODe:CODE 10"

CALL:SSCell:HSSChannel[1]:CONFig:STATe

Function

Sets/queries the secondary serving cell HSDPA/HSPA HS-SCCH 1 Channel Configuration State.

You are not allowed to set the channel configuration to Off when all other HS-SCCH channels' configuration state is Off.

For DC-HSDPA, the number of configured HS-SCCHs across both cells should be 6 or less.

This setting can only be changed when the call status is Idle.

Setting

Range: 1|ON|0|OFF

Query

0|1

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSChannel:CONFig:STATe On"

CALL:SSCell:HSSChannel2:CONFig:STATe

Function

Sets/queries the secondary serving cell HSDPA/HSPA HS-SCCH 2 Channel Configuration State.

You are not allowed to set the channel configuration to Off when all other HS-SCCH channels' configuration state is Off.

For DC-HSDPA, the number of configured HS-SCCHs across both cells should be 6 or less.

This setting can only be changed when the call status is Idle.

Setting

Range: 1|ON|0|OFF

Query

0|1

*RST Setting

1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSChannel2:CONFig:STATe On"

CALL:SSCell:HSSChannel3:CONFig:STATe

Function

Sets/queries the secondary serving cell HSDPA/HSPA HS-SCCH 3 Channel Configuration State.

You are not allowed to set the channel configuration to Off when all other HS-SCCH channels' configuration state is Off.

For DC-HSDPA, the number of configured HS-SCCHs across both cells should be 6 or less.

This setting can only be changed when the call status is Idle.

Setting

Range: 1|ON|0|OFF

Query

0|1

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSChannel3:CONFig:STATe Off"

CALL:SSCell:HSSChannel4:CONFig:STATe

Function

Sets/queries the secondary serving cell HSDPA/HSPA HS-SCCH 4 Channel Configuration State.

You are not allowed to set the channel configuration to Off when all other HS-SCCH channels' configuration state is Off.

For DC-HSDPA, the number of configured HS-SCCHs across both cells should be 6 or less.

This setting can only be changed when the call status is Idle.

Setting

Range: 1|ON|0|OFF

Query

0|1

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:HSSChannel4:CONFig:STATe Off"

CALL:SSCell:OCNSource:CCODe:CODE

Function

Sets/queries the Secondary Serving Cell HSDPA/HSPA OCNS 1-6 Channelization Code

Setting

Range: 1 to 127

Query

Same as setting

*RST Setting

122, 123, 124, 125, 126, 127

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:OCNSource:CCODe:CODE 122, 123, 124, 125, 126, 127"

CALL:SSCell:OCNSource:CONFig:STATe:HSDPa

Function

Sets/queries the Secondary Serving Cell HSDPA/HSPA OCNS 1-6 Channel Configuration State

You are not allowed to set all the HSDPA/HSPA OCNS 1-6 Channel Configuration State for the Secondary Serving Cell to Off at the same time.

Setting

Range: 1|ON|0|OFF

Query

Range: 0|1

*RST Setting

1 (ON), 1 (ON), 1 (ON), 1 (ON), 1 (ON), 1 (ON)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:OCNSource:CONFig:STATe:HSDPa 1,1,1,1,1,1"

CALL:SSCell:OCNSource:LEVel?

Function

Queries the Calculated UTRA FDD Carrier 2 Composite OCNS Level.

Setting

Range: -20.00 to 0, -9.9E37 (NAN)

Query

Same as setting

*RST Setting

-9.9E37

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:OCNSource:LEVel?"

CALL:SSCell:OCNSource:POFFset:HSDPa

Function

Sets/queries the Secondary Serving Cell HSDPA/HSPA OCNS 1-6 Power Offset

Setting

Range: -20.00 to 0 dB

Query

Same as setting

*RST Setting

0, -2, -2, -4, -1, -3

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:OCNSource:POFFset:HSDPa 0, -2, -2, -4, -1, -3"

CALL:SSCell:OCNSource:STATe?

Function

Queries the Calculated UTRA FDD Carrier 2 Composite OCNS State.

Setting

Range: 1|ON|0|OFF

Query

Same as setting

*RST Setting

0 (Off)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:OCNSource:STATe?"

CALL:SSCell:PICHannel:CCODe:CODE

Function

Sets/queries the PICH (SF 256) channelization code for the secondary serving cell. Codes 0 & 1 are not allowed as they are reserved for the CPICH and SCH/PCCPCH respectively.

This setting cannot be changed while the Operating Mode is "Active Cell".

Setting

Range: 2 - 255

Query

Same as setting

*RST Setting

16

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:PICHannel:CCODe:CODE 16"

CALL:SSCell:POWer[:SAMPLitude]

Function

Sets/queries the absolute power of the secondary serving cell.

The setting form of this command performs two functions: (1) sets the desired cell power level and (2) turns the cell power on/off state to on.

This setting is also used in FDD Test Operating Mode.

This setting is used for DC-HSDPA operation.

Setting

Range: -165dBm/3.84MHz to +37dBm/3.84MHz

The actual power range is defined by adding the value of the associated Amplitude Offset to the range in the table. E.g. if Amplitude Offset is -3.5 dB, then the power range will be -118.5dBm/3.84 MHz to -16.5dBm/3.84MHz.

Query

Same as setting

*RST Setting

-75.0 dBm/3.84 MHz

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:POWer -25.7"

CALL:SSCell:POWer:AMPLitude

Function

Sets/queries the absolute power of the secondary serving cell.

This setting is also used in FDD Test Operating Mode.

This setting is used for DC-HSDPA operation.

Setting

Range: -165dBm/3.84MHz to +37dBm/3.84MHz

The actual power range is defined by adding the value of the associated Amplitude Offset to the range in the table. E.g. if Amplitude Offset is -3.5 dB, then the power range will be -118.5dBm/3.84 MHz to -16.5dBm/3.84MHz.

Query

Same as setting

*RST Setting

-75.0 dBm/3.84 MHz

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:POWer:AMPLitude -25.7"

CALL:SSCell:POWer:STATe

Function

Sets/queries the state of the secondary serving cell power.

This parameter only affects the power of the DC-HSDPA secondary serving cell when a DC-HSDPA call is established. Setting this parameter to On/Off will not tell the UE to enable/disable DC-DSDPA operation.

This setting is also used in FDD Test Operating Mode.

Setting

Range: 1|ON|0|OFF

Query

1|0

*RST Setting

ON

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:POWer:STATe On"

CALL:SSCell:SCODe[:PRIMary]

Function

Sets/queries the primary scrambling code of the secondary serving cell. This setting can only be changed when Call/Data Status is Idle .The actual code number for the primary scrambling code is 16 times the value of this setting. See 3GPP TS 25.213, "Spreading and Modulation (FDD)" for a complete description of the relationship between this index and the actual spreading code number.

Setting

Range: 0 to 511

Increment: 1

Query

Same as setting

*RST Setting

2

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:SCODe 2"

CALL:SSCell:TOTal:POWer[:AMPLitude]?

Function

Queries the desired total RF power of the secondary serving cell power for the current system type.

Query

Range: -165 dBm/3.84 MHz to +42 dBm/3.84 MHz or 9.91 E+37 (NAN)

Resolution: 0.01 dBm/3.84 MHz

The actual power range is shown taking the Amplitude Offset into consideration. E.g. if Amplitude Offset is -3.5 dB, then the power range will be -118.5dBm/3.84 MHz to -16.5dBm/3.84MHz.

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:TOTal:POWer?"

CALL:SSCell:TOTal:POWer:STATe?

Function

Queries the total RF power state of the secondary serving cell power for the current system type.

Query

Range: 1|0

*RST Setting

1 (on)

Requirements

Test Application Revision: A.19.00 and above, with the required feature license
Lab Application Revision H.01.00 and above

Programming Example

 
OUTPUT 714;"CALL:SSCell:TOTal:POWer:STATe?"