Test Equipment for 1.85 mm Connector PNA Microwave Network Analyzers

This section lists the required test equipment for the following the PNA series network analyzers:

 

Test Equipment1

Recommended Model

Alternate Model

Sources

Signal Generator

Only used with the U3020A-D01 test set in the Dynamic Accuracy test.

E8257D

N5182B
E8257C
E8267C/D

Counters

Frequency Counter

53230A Option 115

53131A/32A Options 010 or 012, 1242

53150A/51A/52A Option 001

53181A Options 010, 1242

Meters and Power Sensors (See below for calibration service ordering guidelines.)

Power Meter 3
(PowerMeter1)

N1914A4

E4419A/B
EPM-442A4
N1912A

Power Sensor 5

N8482A 3, 6 

8482A

Power Sensor 5

N8487A 3, 6 

8487A

Power Sensor

N8488A

V8486A with the V281A waveguide adapter 10

(comes with Option H67) 9

Attenuators

10 dB Fixed 8

2 required if Option 014, otherwise only one required

8490G Opt. 010

Anritsu 41V-10

20 dB Fixed 8

2 required

8490G Opt. 020

Anritsu 41V-20

Miscellaneous Devices, Test Sets, and Kits

Mechanical or ECal Calibration Kit

85058B mechanical
or
N4694-60001 ECal
(order as N4694A-M0F)

none

Dynamic Accuracy Test Set

U3020A-D01

Z5623A Option H01

Short 1.85 mm (f)11
(SHORT 1)

2 required

May come from 85058B cal kit

Use adapter with male short

85058-60105

None

Load 1.85 mm (f)11

2 required

May come from 85058B cal kit

Use adapter with male short

85058-60112

85058-60124

Load 3.5 mm (m)

May come from 85058B cal kit

909D

902C

USB Mouse

Any

Any

USB Keyboard

Any

Any

Cables

MW, 2.4 mm (f) to 2.4 mm (f)

85133-60002
or
85133-60016

N4697-60100

MW, 1.85 mm (f) to 1.85 mm (f)

N4697-60100

None

1.85 mm (f) to 1.85 mm (m)

Comes with compression test set

2 required

8121-2064

None

2.4 mm (f) to 2.4 mm (m)

Comes with compression test set

2 required

8121-2065

None

GPIB

10833A/B/C/D

Equivalent

Power sensor cable

2 required

11730A

Equivalent

Bypass cable

For Option 014 units only

E8361-60005

None

Adapters

1.85 mm (m) to 1.85 mm (m)

May come from 85058B cal kit

83058-60113

None

2.4 mm (f) to 2.4 mm (f)

Part of cal kit

11900B

85058-60114

(1.85 version from cal kit)

2.4 mm (m) to Type-N (m)

2 required

11903A

Equivalent

Type-N (f) to BNC (m)

1250-1477

Equivalent

2.4 mm (m) to 3.5 (m)

11901A

Equivalent

2.4 mm (f) to Type-N (m)

11903D

08487-60001

2.4 mm (f) to Type-N (f)

11903B

Equivalent

2.4 mm (m) to Type-N (f)

11903C

Equivalent

SMA (f) to BNC (m)

1250-2015

Equivalent

Static Safety Equipment

Adjustable antistatic wrist strap

9300-1367

None

Antistatic wrist strap grounding cord (5 foot)

9300-0980

None

Static control table mat and earth ground wire

9300-0797

None

Optional Equipment

Frequency Reference7

Frequency Counter internal

"house standard"

BNC RF Cable 7

Any

Any

Table Footnotes

  1. Except where noted, one each is the required quantity for all equipment.
  2. A 10 MHz external frequency reference may be used in place of Option 010 or 001.
  3. It is required that your power meter have a compatible firmware version to use the N848xA power sensors. Below are the minimum firmware requirements for each power meter supported by this application:

    • EPM-442A, E4419A, E4419B — Firmware A2.09.01 and above
    • N1912A — Firmware A.05.02 above
  4. Power meters with firmware version A2.03.00 (or A1.03.00) may not operate properly with the new test software. Upgrade the power meter firmware if necessary.

    N1914A power meters with serial number prefix prior to MY53040007 require Service Note N1914A-07 which fixes a power supply ground loop. The ground loop injects noise into the measurement circuits which can cause unstable measurements at low (–30 dBm) power levels

  5. A current calibration report with data is required for each power sensor used. The user must enter this data for each power sensor.
  6. Both Standard (STD) and Cal Factor Option (CFT) versions of the N848xA power sensors are supported in this application. The correct version needs to be selected when configuring the test station in the Administration section of TME.
  7. The frequency reference is optional. It may be a 10 MHz "house standard", or it may be the internal frequency reference time-base of the counter used for the frequency measurement. See the Frequency Reference Explanation section in the Frequency Accuracy Test for details. The BNC cable is only required if a 10 MHz external reference is used.
  8. Fixed attenuators are not needed if there is a step attenuator in the PNA.

  9. Option H67 is an ANSI Z540-1 calibration for the V8486A power sensor with a V281A connector adapter (connected together).
  10. Combination of the V8486A power sensor and V281A adapter must be calibrated as a single unit—do not separate. For more information, click here.Closed
  11. When configuring equipment assignments for a test station, these parts may be configured as PartsInPlaceOfCalKit for CalkitTwoPort1 when an 85058B/E cal kit is not used.

Guidelines for Ordering Power-Sensor Calibration Service

Use the following guidelines to ensure power sensors used by the N7843A application receive the correct calibration service:

  1. Order Option 1A7 when ordering new power sensors that will be used as working standards in N7800A TME calibrations.

  2. Order the Keysight Cal + Measurement Uncertainty + GB calibration service for power sensor recalibration.

  3. When ordering periodic calibration for instruments used as lab standards in the N7800A software, we recommend using “Keysight calibration + uncertainties” for power sensors, and “Keysight calibration + uncertainties + guardbanding” for all other items (please visit Selecting the Right Calibration Services). The N7800A software incorporates the ISO GUM Uncertainty in point-to-point uncertainty calculations. The special “H-series” calibration options in this table provide lower measurement uncertainties through use of direct comparison to devices directly characterized by NPL or NIST (or another NMI). Please order Option H99 to get electronic data for easy import into the N7800A (avoids manual entry). The overall resulting N7800A measurement uncertainties then reflects these lower device uncertainties. The equipment requirements of each N7800A calibration application are summarized in the specials calibration matrix which can be found at the Recommended Lab Standards and Special Cal Options website.

  4. The Keysight Option H99 is a special option for the Roseville Service Center [only] which provides electronic calibration data available as a .csv file in TME N7800A format for all power sensors used with the N7800A TME application. N7800A TME is SSU calibration software used across Keysight SSU and by many self-maintainers. H99 must be requested from the Roseville SSU upon re-calibration only of any previously purchased power sensor and not for new purchases of BID Power Sensors. This option may be ordered in addition to any other required or requested Std Lab Calibration option (must also be ordered in addition to H99). Option H99 is recommended but not required.