The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again. |
This test checks the dynamic accuracy at 1.195 GHz for each receiver port using the Keysight Z5623A (with Option H01) dynamic accuracy test set. The 1.195 GHz source signal of the DUT goes through the test set's first attenuator variable from 0 dB to 11 dB in 1 dB steps and second attenuator variable up to 100 dB in 10 dB steps.
The test procedure checks the measured receiver-input power for each 10 dB increment in the range from 0 dBm to –60 dBm for 35 dB reference/test attenuator, and from –20 dBm to –80 dBm for 0 dB reference/test attenuator. The ratios of the measured receiver-input powers to the reference input level of –10 dBm (for 35 dB reference/test attenuator) or –30 dBm (for 0 dB reference/test attenuator) are calculated and compared to the attenuation values of the test set.
The 1.195 GHz signal flow is split with a power splitter present between the two attenuators in the test set and measured with an external power meter to calibrate the attenuated power level.
|
Test Equipment |
Recommended Model Number1 |
|---|---|
|
Power meter |
N1914A |
|
Power sensor |
N8482A |
|
6 dB fixed attenuator |
8491B Option 006 |
|
Dynamic accuracy test set |
Z5623A-H01 |
|