Dynamic Accuracy — 1 MHz

(M9376A)

The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

This test checks the dynamic accuracy at 1 MHz. The dynamic accuracy is measured at the receiver port at 1 MHz from –60 dBm to 0 dBm in 10 dB steps using a step attenuator.

Required Test Equipment

Test Equipment

Recommended Model Number1

Power meter

N1914A

Power sensor

E9304A-H18

10 dB step attenuator

8496G Option H52

Attenuator switch driver

11713B

  1. Refer to the main equipment list for alternative models.

Connection Setups

Power sensor calibration

Source calibration

Dynamic Accuracy