The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again. |
This test checks the dynamic accuracy at 50 MHz. The dynamic accuracy is measured at the test input (T In) and reference input (R In) ports at 50 MHz, from –60 dBm to 0 dBm for 35 dB reference/test attenuator, and from –80 dBm to –20 dBm for 0 dB reference/test attenuator, in 10 dB steps using a step attenuator.
|
Test Equipment |
Recommended Model Number1 |
|---|---|
|
Power meter |
N1914A |
|
Power sensor |
E9304A-H19 |
|
10 dB step attenuator |
8496G |
|
Attenuator switch driver |
11713B |
|