Maximum Leveled Power

(M9376A)

  • The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

  • This test requires the dividers be disconnected from the receiver modules. At any time that a reference signal is not connected to both of the receiver modules, and the M9485A IVI Driver is initialized, the firmware will generate an FPGA error message that can be dismissed.

A source with specified power is connected to the Source-In port of the M9376A receiver module. The test measures the output power at varying test frequencies, and verifies the maximum power listed in the Max Leveled Power specification under the Source output of the M9376A receiver module.

Required Test Equipment

Test Equipment

Recommended Model Number1

Power meter

N1914A

Power sensor

E9304A-H18

Network analyzer

E5071C2

Adapter for receiver module

M9376-65001

  1. Refer to the main equipment list for alternative models.
  2. The frequency of the E5071C network analyzer should be greater than or equal to the maximum frequency of the UUT.

Connection Setups

Power sensor calibration

Source calibration

Maximum leveled power measurement