The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again. |
This test checks the noise floor of the test port on the M9376A receiver module. Since absolute measurement accuracy is not a hard specification, the noise floor is determined by first obtaining a correction coefficient between an absolute measurement value and a power meter reading. This reading is then applied to an absolute measurement performed on the DUT while the test port is terminated with an Open and then a Load to calculate the noise floor.
|
Test Equipment |
Recommended Model Number1 |
|---|---|
|
Power meter |
N1914A |
|
Power sensor |
E9304A-H18 |
|
Open |
Any |
|
Load |
Any |
|