Noise Floor

(M9376A)

The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

This test checks the noise floor of the test port on the M9376A receiver module. Since absolute measurement accuracy is not a hard specification, the noise floor is determined by first obtaining a correction coefficient between an absolute measurement value and a power meter reading. This reading is then applied to an absolute measurement performed on the DUT while the test port is terminated with an Open and then a Load to calculate the noise floor.

Required Test Equipment

Test Equipment

Recommended Model Number1

Power meter

N1914A

Power sensor

E9304A-H18

Open

Any
(can come from the 85052D cal kit)

Load

Any
(can come from the 85052D cal kit)

  1. Refer to the main equipment list for alternative models.

Connection Setups

Power sensor calibration

Source calibration setup

Noise Floor measurement setup with an Open

Noise Floor measurement setup with a Load