The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again. |
This test checks the noise floor of the test input (T In) and reference input (R In) ports on the M9377A direct access receiver module. Since the absolute measurement accuracy is not a warranted specification, the absolute measurement is made using an E5071C network analyzer which the power level is calibrated by using a power meter and power sensor. The absolute measurement correction will be applied to the raw rms noise measurement performed on the M9377A while the receiver ports are terminated with 50 Ohm loads to calculate the noise floor.
|
Test Equipment |
Recommended Model Number1 |
|---|---|
|
Network analyzer |
E5071C2 |
|
Power meter |
N1914A |
|
Power sensor |
E9304A-H19 |
|
Load |
Any |
|