Receiver Compression

(M9376A)

  • The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

  • A fail result in this test implies that Dynamic Accuracy specification cannot be guaranteed in the compression range of the receiver.

This test measures the compression at the specified maximum power level for the receivers. The test process measures two power deltas, one at the top of the receiver input power range, and another at an expected linear part of the input power range. The difference between the deltas represents the compression at the top of the power range. A negative test result indicates compression while a positive result indicates expansion.

Required Test Equipment

Test Equipment

Recommended Model Number1

Power meter

N1914A

Power sensor

E9304A-H18

10 dB step attenuator

8496G Option H52

Attenuator switch driver

11713B

  1. Refer to the main equipment list for alternative models.

Connection Setups

Power sensor calibration

Source calibration

Receiver compression