Return Loss

(M9377A)

The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

This test measures the return loss of the source output port, and the return loss of the receiver test input (T In) and reference input (R In) ports of the M9377A. The return loss of the M9377A is measured using an E5071C after the E5071C has performed a one-port calibration.

Ensure that the adapter and cable remain connected to the PORT 1 of the network analyzer until the Return Loss test is completed. Otherwise, the Network Analyzer Calibration will need to be repeated.

Required Test Equipment

Test Equipment

Recommended Model Number1

Network analyzer

E5071C2

Calibration kit

85052D, mechanical
85093-60010 (85093C Option 00F), ECal

  1. Refer to the main equipment list for alternative models.
  2. The frequency of the E5071C network analyzer should be greater than or equal to the maximum frequency of the DUT.

Connection Setups

Network analyzer calibration, mechanical calibration kit

Network analyzer calibration, ECal module

Return loss measurement, Src Out Port

 

Return loss measurement, T In Port

 

Return loss measurement, R In Port