Uncorrected System Performance

(M9376A)

The Power Correction Calibration must be performed prior to running this test. The results are valid for 48 hours after such time it will need to be performed again.

This test checks the directivity, source match, and load match which are the key S-parameter measurement hardware characteristics.

These characteristics are tested using a calibration kit with segment sweep points for a specified frequency range. After a port calibration is performed, the test will output the directivity, source match, and load match data to the report.

Required Test Equipment

Test Equipment

Recommended Model Number1

Calibration kit

85052D

  1. Refer to the main equipment list for alternative models.

Connection Setups

To expand all, click the Expand All button in the toolbar (webhelp only).

Directivity, source match, and load match

ClosedClick here for images.