Calibration Mode
Calibration Mode affects when VSA indicates a calibration needed. If Auto Calibration is enabled, a quick calibration is run on the sections that need to be recalibrated (aligned) based on the selected Calibration Mode.
- Precision: The instrument will satisfy the specifications. Calibrates less often than Classic. Affected by time and temperature.
- Normal: The instrument will perform well but might not satisfy the specifications. Calibrates less often than Precision. Affected by time (less so than Precision) and temperature.
- Light: The instrument will perform well but might not satisfy the specifications. Calibrates less often than Normal. Affected by temperature.
- Classic: The instrument will satisfy the specifications. Calibrates the most often. Affected by time and temperature.
See Also