CALCulate:SA:MARKer commands

Controls the marker settings used in the SA application.

These commands are Superseded by the CALCulate:MEASure:SA commands.

 

CALCulate:SA:MARKer:

BDENsity

       | ACPR

           | [:STATe]

       | BW

       | DATA?

       | EQSPan

       | NOISe

           | [:STATe]

       | NPR

           | [:STATe]

       | POWer

           | BW

           | [:STATe]

       | TONE

           | BW

           | [:STATe]

           | TSPacing

BPOWer

       | DATA?

       | SPAN

       | [:STATe]

OCCBand

        | CENTer?

        | PERCent

        | POWer?

        | SPAN?

        | [:STATe]

 

Click on a keyword to view the command details.

See Also

Critical Note: CALCulate commands act on the selected measurement. You can select one measurement for each channel using Calc:Par:MNUM or Calc:Par:Select. Learn more.

If you use Calc:Par:Cat? to return the list of current measurements, the returned string will be similar to ‘CH1_B_1,B’.

To select this measurement as a parameter for SA, you need to send Calc:Par:Sel ‘CH1_B_1’.

Moreover, most of the following commands will return ‘+202, “Parameter not valid” if they are called with marker number n, and marker number n is not currently turned ON.

Important: Learn about programming the reference marker.

Note: For all band power marker family (this includes BNOise, BPOWer, OCCBand markers) and when measuring wideband repetitive modulated signals, there is basically 2 approaches to get good measurements:

- Either run the coherent mode of SA if the modulated test signal repetition period or tone spacing is known and a frequency reference connection is made between the signal source and the PNA (usually the 10 MHz ref signal BNC). See SA Setup Coherence.

- Or make use of a RBW that is equal or smaller than 1/20 of the modulated test signal tone spacing (rule of thumb). The test signal tone spacing is 1/ the test signal duration. For example, if the ARB file that generates the test signal replays each 100 microseconds, it means the tone spacing is 10 kHz. So, we recommend to set the RBW to 500 Hz or below.


CALCulate<ch>:SA:MARKer<n>:BDENsity:ACPR[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Sets and reads the ACPR density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density marker OFF.

1 - ON - Turn band density marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC2:SA:MARK3:BDEN:ACPR:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:ACPR[:STATe]?

Return Type

Boolean

Default

0


CALCulate<ch>:SA:MARKer<n>:BDENsity:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:SA:MARK:BDEN:BW 1e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:BW?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:SA:MARKer<n>:BDENsity:DATA?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the band density level in dBm/Hz from the band density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that SA channel. If unspecified, <ch> is set to 1.

<n>

Band density marker number. If unspecified, <n> is set to 1.

Examples

CALC:SA:MARK:BDEN:DATA?

calculate2:sa:marker2:bdensity:data?

Return Type

Numeric

Default

Not applicable


CALCulate<ch>:SA:MARKer<n>:BDENsity:EQSPan <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the frequency span used by Power Density to normalize the power.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a span.

Examples

CALC:SA:MARK:BDEN:EQSPan 1e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:EQSPan?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:SA:MARKer<n>:BDENsity:NOISe[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band density noise marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density noise marker OFF.

1 - ON - Turn band density noise marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:SA:MARK:BDEN:NOIS:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:NOISe?

Return Type

Boolean

Default

0


CALCulate<ch>:SA:MARKer<n>:BDENsity:NPR[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Sets and reads the NPR density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density marker OFF.

1 - ON - Turn band density marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC2:SA:MARK3:BDEN:NPR:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:NPR[:STATe]?

Return Type

Boolean

Default

0


CALCulate<ch>:SA:MARKer<n>:BDENsity:POWer:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band power density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:SA:MARK:BDEN:POW:BW 1e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:POWer:BW?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:SA:MARKer<n>:BDENsity:POWer[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band power density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band power density marker OFF.

1 - ON - Turn band power density marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:SA:MARK:BDEN:POW:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:POWer?

Return Type

Boolean

Default

0


CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band tone density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:SA:MARK:BDEN:TONE:BW 1e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE:BW?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band tone density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band tone density marker OFF.

1 - ON - Turn band tone density marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:SA:MARK:BDEN:TONE:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE?

Return Type

Boolean

Default

0


CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE:TSPacing <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the spacing of the band tone density marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a spacing value.

Examples

CALC:SA:MARK:BDEN:TONE:TSP 100e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BDENsity:TONE:TSPacing?

Return Type

Numeric

Default

100 MHz


CALCulate<ch>:SA:MARKer<n>:BPOWer:DATA?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the band power level from the band power marker.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that SA channel. If unspecified, <ch> is set to 1.

<n>

Band power marker number. If unspecified, <n> is set to 1.

Examples

CALC:SA:MARK:BPOW:DATA?

calculate2:sa:marker2:bpower:data?

Default

Not applicable


CALCulate<ch>:SA:MARKer<n>:BPOWer:SPAN <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the frequency span of  the band power marker. This area is marked by two vertical dotted lines on the screen and the marker's y-axis value is set to the measured power value. Noise and power on the same marker share the same span.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a frequency span within the frequency range of the analyzer.

Examples

CALC:SA:MARK:BPOW:SPAN 1e6

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BPOWer:SPAN?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:SA:MARKer<n>:BPOWer[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band power marker. This command makes a band power marker from a generic marker. The generic marker must first be created using: CALC:MARK:STATe

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band power marker OFF.

1 - ON - Turn band power marker ON.

Examples

'Select the measurement

CALC2:PAR:SEL "M2SA_CH2_A"

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band power marker

CALC:SA:MARK:BPOW:STAT 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:BPOWer?

Return Type

Boolean

Default

0

CALCulate<ch>:SA:MARKer<n>:OCCBand:CENTer?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth center frequency.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that SA channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

Examples

CALC:SA:MARK:OCCB:CENT?

calculate2:sa:marker2:occband:center?

Default

Not applicable


CALCulate<ch>:SA:MARKer<n>:OCCBand:PERCent <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and returns the percentage of the band power to search for.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Percentage value.

Examples

CALC:SA:MARK:OCCB:PERC 99

calculate2:sa:marker2:occband:percent 99

Query Syntax

CALCulate<ch>:SA:MARKer<n>:OCCBand:PERCent?

Return Type

Numeric

Default

99.0


CALCulate<ch>:SA:MARKer<n>:OCCBand:POWer?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth power.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that SA channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

Examples

CALC:SA:MARK:OCCB:POW?

calculate2:sa:marker2:occband:power?

Default

Not applicable


CALCulate<ch>:SA:MARKer<n>:OCCBand:SPAN?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth span.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that SA channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

Examples

CALC:SA:MARK:OCCB:SPAN?

calculate2:sa:marker2:occband:span?

Default

Not applicable


CALCulate<ch>:SA:MARKer<n>:OCCBand[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and returns the occupied bandwidth on/off state.

See Critical Note

Parameters

 

<ch>

Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <ch> is set to 1.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turns occupied bandwidth OFF.

1 - ON - Turns occupied bandwidth ON.

Examples

CALC:SA:MARK:OCCB:STAT 1

calculate2:sa:marker2:occband:state 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:OCCBand[:STATe]?

Return Type

Boolean

Default

0