:CORR1:COLL:DC

Syntax

[:SENSe]:CORRection1:COLLect[:ACQuire]:DC {STAN1|STAN2|STAN3}

Description

Obtains the data for calculating the calibration coefficients to use during Rdc measurement performed with the specified standard. The data is obtained for only the active table. (No query)

:INIT:CONT OFF
'
'
:CORR1:COLL:DC STAN1
:INIT
*OPC?

Variable

Parameter

Param

Range

Select from the following:

STAN1: OPEN standard

STAN2: SHORT standard

STAN3:LOAD standard

Related objects

:CORR1:COLL

:CORR1:COLL:RF

Equivalent key

Cal > OPEN|SHORT|LOAD > DC