Property (Read-Write)
SCPI.SENSe.PS(Ch).FREQuency.SPAN = Value
Value = SCPI.SENSe.PS(Ch).FREQuency.SPAN
This command sets/gets span frequency, for the selected channel Ch.
Parameter |
Value |
Description |
Span frequency |
Data Type |
Double precision floating point type (Double) |
Range |
200m ~ 25M |
Preset Value |
999k |
Unit |
Hz |
Resolution |
200m |
Preset Effect |
Yes |
Save-Recall Effect |
Yes |
Error |
- |
Dim Var as Double
Var= 10e6
SCPI.SENSe.PS1.FREQuency.SPAN = Var
Var = SCPI.SENSe.PS1.FREQuency.SPAN
SCPI.SENSe.PS(Ch).FREQuency.CENTer
SCPI.SENSe.PS(Ch).FREQuency.STARt
SCPI.SENSe.PS(Ch).FREQuency.STOP
PS Menu: Start/Center > Span
PS Menu: Stop/Span > Span
PS Menu: Marker To > Marker -> Span
:SENSe:PS[1-1]:FREQuency:SPAN {200m ~ 25M }
:SENSe:PS[1-1]:FREQuency:SPAN?
{200m ~ 25M } <newline><^END>
10 OUTPUT 717;":SENSe:PS1:FREQuency:SPAN 10e6"
20 OUTPUT 717;":SENSe:PS1:FREQuency:SPAN ?"
30 ENTER 717;A