Define the settings for Transient measurement class.
SENSe:TR | STYPe | NARRow[1-2] | BANDwidth | VIDeo | [:VALue] | AUTO | FREQuency | PORT | TIME | POWer | INPut | AUTO | CORRection | DATA | LOAD | STATe | LEVel | TIME | COUPling | WIDE | FREQuency | BANDwidth | VIDeo | [:VALUe] | AUTO | MINimum | PORT
|
|
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and read the transition mode, Wide-Narrow or Narrow-Narrow. |
|
Parameters |
|
<type> |
Choose from: WN - Wide-Narrow measurement mode NN - Narrow-Narrow measurement mode |
Examples |
SENS:TR:STYP NN sense:tr:stype WN |
Query Syntax |
SENSe:TR:STYpe? |
Return Type |
Character |
Default |
WN |
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and read the Video Band Width (VBW) for Narrow1 or Narrow2. |
|
Parameters |
|
<value> |
Video Band Width for Narrow band |
Examples |
SENS:TR:NARR1:BAND:VID sense:tr:narrow2:bandwidth:video |
Query Syntax |
SENSe:TR:NARRow[1-2]:BANDwidth:VIDeo[:VALUe]? |
Return Type |
Numeric |
Default |
1.7M |
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and read the Video Band Width (VBW) AUTO mode for Narrow1 or Narrow2. |
|
Parameters |
|
<bool> |
Choose from: 0 - OFF - Disable the auto VBW 1 - ON - Enable the auto VBW |
Examples |
SENS:TR:NARR1:BAND:VID:AUTO ON sense:tr:narrow2:bandwidth:video:auto 0 |
Query Syntax |
SENSe:TR:NARRow[1-2]:BANDwidth:VIDeo:AUTO? |
Return Type |
Numeric |
Default |
ON |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the center frequency for Narrow1 or Narrow2. |
|
Parameters |
|
<value> |
Center frequency for Narrow band |
Examples |
SENS:TR:NARR1:FREQ:CENT sense:tr:narrow2:frequency:center |
Query Syntax |
SENSe:TR:NARRow[1-2]:FREQuency:CENTer? |
Return Type |
Numeric |
Default |
1G |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the input port for narrow band 1 and 2. Only one port can be selected in channel. Hence, the same value is set on Narrow 1,2 and Wide. The value is overwritten. |
|
Parameters |
|
<value> |
Input port. |
Examples |
SENS:TR:NARR1:PORT 2 sense:tr:narrow2:frequency:port 1 |
Query Syntax |
SENSe:TR:NARRow[1-2]:PORT? |
Return Type |
Numeric |
Default |
1 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the frequency span of narrow band 1/2. |
|
Parameters |
|
<freqrange> |
Choose from: R3K - 3.125 kHz Frequency Span R25K - 25 kHz Frequency Span R0_3 - 312.5 kHz Frequency Span R2_5 - 2.5 MHz Frequency Span R10 -10 MHz Frequency Span R30 - 30 MHz Frequency Span R80 - 80 MHz Frequency Span R160 - 160 MHz Frequency Span R320 - 320 MHz Frequency Span |
Examples |
SENS:TR:NARR1:FREQ:RANG R3K sense:tr:narrow2:frequency:range R25K |
Query Syntax |
SENSe:TR:NARRow[1-2]:FREQuency:RANGe? |
Return Type |
String |
Default |
R30 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the offset time at the reference point for the narrow band 1 or 2. |
|
Parameters |
|
<value> |
Offset time |
Examples |
SENS:TR:NARR[1-2]:TIME:OFFS sense:tr:narrow[1-2]:time:offset |
Query Syntax |
SENSe:TR:NARRow[1-2]:TIME:OFFSet? |
Return Type |
Numeric |
Default |
0 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the position at the reference point of the narrow band 1 or 2. |
|
Parameters |
|
<type> |
Choose from: LEFT CENTer RIGHt |
Examples |
SENS:TR:NARR1:TIME:REF LEFT sense:tr:narrow2:time:reference right |
Query Syntax |
SENSe:TR:NARRow[1-2]:TIME:REFerence? |
Return Type |
Character |
Default |
CENT |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the time span of the narrow band 1 or 2. |
|
Parameters |
|
<value> |
Time span in sec |
Examples |
SENS:TR:NARR1:TIME:SPAN 0.01 sense:tr:narrow2:time:span 0.01 |
Query Syntax |
SENSe:TR:NARRow[1-2]:TIME:SPAN? |
Return Type |
Numeric |
Default |
|
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the receiver attenuation value |
|
Parameters |
|
<value> |
Attenuation level in dB |
Examples |
SENS:TR:POW:INP:ATT 20 sense:tr:power:input:attenuation 15 |
Query Syntax |
SENSe:TR:POWer:INPut:ATTenuation? |
Return Type |
Numeric |
Default |
10 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the on/off of receiver attenuation auto function |
|
Parameters |
|
<bool> |
Choose from: 0 - OFF - Disable the auto attenuation 1 - ON - Enable the auto attenuation |
Examples |
SENS:TR:POW:INP:ATT:AUTO on sense:tr:power:input:attenuation:auto 1 |
Query Syntax |
SENSe:TR:POWer:INPut:ATTenuation:AUTO? |
Return Type |
Boolean |
Default |
OFF |
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and returns the power correlation data. The correction data can be loaded from the csv file using SENS:TR:POW:INP:CORR:LOAD. See Data Map and Correction. No GUI for this command. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<array> |
A set of frequency and power correction data. The separator is a comma. |
Examples |
SENS:TR:POW:INP:CORR:DATA
10E+6,0.04,100E+6,0.06,200E+6,0.07,300E+6,0.06 |
Query Syntax |
SENSe<ch>:TR:POWer:INPut:CORRelation:DATA? |
Return Type |
Numeric Array |
Default |
0,0 |
Applicable Models: E505xA with opt S96301xB (Write Only) Sets and returns the power correlation data from correction definition file. The format of data is a frequency value and a correction value for the frequency with a comma (,) separator. See Data Map and Correction. No GUI for this command. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<file> |
file name |
Examples |
SENS:TR:POW:INP:CORR:LOAD
"C:\CorrData.csv" |
Query Syntax |
N/A |
Return Type |
N/A |
Default |
N/A |
Applicable Models: E505xA with opt S96301xB (Write Only) Enable and disable the power correlation. The correction data can be set by SENS:TR:POW:INP:CORR:LOAD (csv file) or SENS:TR:POW:INP:CORR:DATA. See Data Map and Correction. No GUI for this command. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<bool> |
Choose from: 0 - OFF - Disable the power correction. 1 - ON - Enable the power correction. |
Examples |
SENS:TR:POW:INP:CORR:STAT
ON |
Query Syntax |
SENSe<ch>:TR:POWer:INPut:CORRelation:STATe? |
Return Type |
Boolean |
Default |
OFF |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the maximum power input level. |
|
Parameters |
|
<value> |
Maximum power input level in dBn |
Examples |
SENS:TR:POW:INP:LEV:MAX sense:tr:power:input:level:maximum |
Query Syntax |
SENSe:TR:POWer:INPut:LEVel:MAXimum? |
Return Type |
Numeric |
Default |
|
Applicable Models: E505xA with opt S96301xB (Read-Write) Enable and disable the time coupling for two measurements (Wide-Narrow or Narrow1-Narrow2). |
|
Parameters |
|
<Boolean> |
Choose from: OFF / 0 ON /1 |
Examples |
SENS:TR:TIME:COUP[:STAT] OFF sense:tr:time:coupling[:state] on |
Query Syntax |
SENSe:TR:TIME:COUPling[:STATe]? |
Return Type |
Character |
Default |
ON |
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and read the Video Band Width (VBW) for Wide Band. |
|
Parameters |
|
<value> |
Video Band Width for Narrow band |
Examples |
SENS:TR:WIDE:BAND:VID sense:tr:wide:bandwidth:video |
Query Syntax |
SENSe:TR:WIDE:BANDwidth:VIDeo? |
Return Type |
Numeric |
Default |
250E3 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Sets and read the Video Band Width (VBW) AUTO mode for wide band. |
|
Parameters |
|
<bool> |
Choose from: 0 - OFF - Disable the auto VBW 1 - ON - Enable the auto VBW |
Examples |
SENS:TR:WIDE:BAND:VID:AUTO ON sense:tr:wide:bandwidth:video:auto 0 |
Query Syntax |
SENSe:WIDE:BANDwidth:VIDeo:AUTO? |
Return Type |
Numeric |
Default |
ON |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the maximum frequency when the frequency range is over 8 GHz. |
|
Parameters |
|
<value> |
Maximum frequency in Hz |
Examples |
SENS:TR:WIDE:FREQ:MAX sense:tr:wide:frequency:maximum |
Query Syntax |
SENSe:TR:WIDE:FREQuency:MAXimum? |
Return Type |
Numeric |
Default |
13E9 |
Applicable Models: E505xA with opt S96301xB (Read only) Read the maximum frequency when the frequency range is over 8 GHz. |
|
Parameters |
|
Examples |
SENS:TR:WIDE:FREQ:MIN? sense:tr:wide:frequency:minimum? |
Query Syntax |
SENSe:TR:WIDE:FREQuency:MINimum? |
Return Type |
Numeric |
Default |
6E9 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the frequency range for wide band. |
|
Parameters |
|
<freqrange> |
Choose from: R4 - 40 MHz – 160 MHz Frequency Range R8 - 80 MHz – 320 MHz Frequency Range R16 - 160 MHz – 640 MHz Frequency Range R32 - 320 MHz – 1.28 GHz Frequency Range R64 - 640 MHz – 2.56 GHz Frequency Range R128 - 1.28 GHz – 5.12 GHz Frequency Range R256 - 2.56 GHz – 8 GHz Frequency Range R40M - 1 MHz – 40 MHz Frequency Range R8G - 256 MHz – 8 GHz Frequency Range RMAX - Max.Freq. > 8 GHz |
Examples |
SENS:TR:WIDE:FREQ:RANG R4 sense:tr:wide:frequency:range R40M |
Query Syntax |
SENSe:TR:WIDE:FREQuency:RANGe? |
Return Type |
String |
Default |
R8G |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the input port for wide band. Only one port can be selected in channel. Hence, the same value is set on Narrow 1,2 and Wide. The value is overwritten. |
|
Parameters |
|
<value> |
Input port. |
Examples |
SENS:TR:WIDE:PORT 2 sense:tr:wide:port 1 |
Query Syntax |
SENSe:TR:WIDE:PORT? |
Return Type |
Numeric |
Default |
1 |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the time offset for wide band. |
|
Parameters |
|
<value> |
Time offset in sec |
Examples |
SENS:TR:WIDE:TIME:OFFS 0.01 sense:tr:wide:time:offset 0.02 |
Query Syntax |
SENSe:TR:WIDE:TIME:OFFSet? |
Return Type |
Numeric |
Default |
|
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the position at the reference point of wide band. |
|
Parameters |
|
<type> |
Choose from: LEFT CENTer RIGHt |
Examples |
SENS:TR:WIDE:TIME:REF LEFT sense:tr:wide:time:reference right |
Query Syntax |
SENSe:TR:WIDE:TIME:REFerence? |
Return Type |
Character |
Default |
CENT |
Applicable Models: E505xA with opt S963015B (Read-Write) Set and read the time span of wide band. |
|
Parameters |
|
<value> |
Time span in sec |
Examples |
SENS:TR:WIDE:TIME:SPAN sense:tr:wide:time:span |
Query Syntax |
SENSe:TR:WIDE:TIME:SPAN? |
Return Type |
Numeric |
Default |
|