Preparation for Measurement

Other topics about Overview

Measurement Method

The E5061B allows you to several measurement method according to your DUT. Each method requires the following test configuration.

Measurement using Ports 1 and 2

Method

Required Items

Connection

Port 1

Reflection

16201A Terminal Adapter, Keysight 7 mm type test fixture, Calibration Kit 16195B (or 85031B)

Port 1-2

Series - Through

User-prepared fixture, Cables, 50 Ω precision resistor for calibration at fixture

Port 1-2

Shunt - Through

User-prepared shunt-through fixture, Cables, Calibration kit

Measurement using Gain-Phase Port

Method

Required Items

Connection

Gain-Phase

Series - Through

Keysight 4 terminal-pair type test fixture (Recommendation 16047E for leaded DUTs, 16034E/G/H for SMDs), 50 Ω precision resistor for calibration at fixture

Gain-Phase

Shunt - Through

User-prepared fixture, Cables, Power splitter (11667L or equivalent)