The test set provides a 5 V, 16.5 us pulse at the TRIG OUT rear panel connector for triggering other test equipment from the test set.
This pulse is generated according to the
Trigger Output Setup
setting:
System Frame Clock
When
Trigger Output Setup
is set to
System Frame Clock
, a pulse is generated every 10 ms, in alignment with the first symbol period of each frame.
Measurement Generated
When
Trigger Output Setup
is set to
Measurement Generated
, pulses are generated when certain measurements are initiated or are running, or for other special situations, as outlined below:
Out-of-Synch Handling of Output Power measurement:
When
Out-of-Synch Handling of Output Power
is initiated, the test set generates a trigger pulse at point "A" (see
Out-of-Synch Handling of Output Power Measurement Description
for more details).
Note, when
Trigger Output Setup
is set to
Measurement Generated
and HSDPA is not active, the signal present at the
TRIG OUT
connector is undefined when any measurements other than
Out-of-Synch Handling of Output Power
is triggered or running.
HSDPA ( lab application or feature-licensed test application only ):
When HSDPA is active, a trigger pulse is generated every time the test set sends data using the HARQ0 process.
When on an HSDPA or HSPA RB Test Mode connection with
RB Test HS-DSCH Configuration Type
set to
FRC
, downlink HARQ0 subframes occur every 12 ms. Thus, the downlink HARQ0 rear panel trigger occurs every 12 ms. However, with
HS-DSCH Configuration Type
set to
User Defined
, or while on an HSDPA or HSPA packet data connection, downlink HARQ0 subframes do not necessarily occur every 12 ms (for example, if you are not continuously sending data on a packet data connection, the HARQ0 process is not invoked every 12 ms).
For a definition of when HSDPA is active for active cell operating mode and FDD test operating mode, see HSDPA Activation in Active Cell and HSDPA Activation in FDD Test , respectively.
Note, the rear panel trigger is NOT affected by the
HS-DPCCH Trigger Subframe Alignment
settings present in many measurements (which allow you to move the
measurement
trigger relative to the downlink HARQ0 subframe). When present, the rear panel trigger is always aligned to the start of the downlink HARQ0 subframes.
To access
Trigger Output Setup
from the front panel, press the
CALL SETUP
key, press the
More
key until you reach the
Call Control 2 of 5
menu, and then select
Trig Output Setup
(
F6
).