Last updated: January 16, 2009
This section is only applicable to the lab application.
Out-of-synchronization handling of output power measures the UE's ability to monitor the downlink DPCCH quality and turn its transmitter on or off based on the quality level.
3GPP TS 34.121 5.4.4.2 states: "When the UE estimates the DPCCH quality over the last 160 ms period to be worse than a threshold Qout, the UE shall shut its transmitter off within 40 ms. The UE shall not turn its transmitter on again until the DPCCH quality exceeds an acceptable level Qin. When the UE estimates the DPCCH quality over the last 160 ms period to be better than a threshold Qin, the UE shall again turn its transmitter on within 40 ms."
"The UE shall not shut its transmitter off before point B. The UE shall shut its transmitter off before point C, which is Toff = 200 ms after point B. The UE shall not turn its transmitter on between points C and E. The UE shall turn its transmitter on before point F, which is Ton = 200 ms after point E."
The UE's transmitter is considered to be "off" if the transmitted power is below -55 dBm and "on" if the transmitted power is above -49 dBm.
The
Out-of-Synch Handling of Output Power
measurement allows you to specify the time intervals and DPCCH_Ec/Ior levels between points A and F (see
Out-of-Synch Handling of Output Power Measurement Parameters
). When the measurement is initiated, the test set varies its DPCCH level according to these settings. You must then measure the UE's power response using an external spectrum analyzer to determine if the UE turned its power off and on as required (see
Out-of-Synch Handling of Output Power Measurement Results
).
Before performing the out-of-synch measurement, you must:
Measurement Generated
.The following settings are recommended based on the 3GPP TS 34.121 5.4.4 requirements. Note that some of these parameters must be set before establishing a connection.
Connected DL Channel Levels
as specified by 3GPP TS 34.121 Table E.3.3 and Table 5.4.4.3:
Cell 1 Connected CPICH Level
=
-10.00 dB
.
Cell 1 Connected P-CCPCH/SCH Level
=
-12.00 dB
.
Cell 1 Connected S-CCPCH Level
=
Off
.
Cell 1 Connected PICH Level
=
-15.00 dB
.
Cell 1 Connected DPCH Level
=
-16.60 dB
. (This sets DPCCH_Ec/Ior before point A to -16.6 dB, because PO1 = PO2 = PO3 = 0, which makes DPCH_Ec/Ior = DPCCH_Ec/Ior).See Downlink Channel Levels .
All Up bits
.
Call Parms 1 of 3
menu) to
-61 dBm
.
Generator Info
(
F3
on the
Call Control 2 of 4
menu) and set
AWGN Power
(
F5
) to
-60
dBm.No other measurements can be running when an out-of-synch handling of output power measurement is performed. If any measurements are running when an out-of-synch handling of output power measurement is initiated, they are closed and a message is displayed to indicate that they were closed. Likewise, if any other measurements are initiated while an out-of-synch handling of output power measurement is running, the out-of-synch handling of output power measurement is closed and a message is displayed.
The out-of-synch handling of output power measurement can be performed while in FDD Test Operating Mode , however you must configure the UE such that its output power will respond appropriately to the changes in the test set's DPCCH_Ec/Ior.
See 3GPP TS 34.121 Figure 5.4.4.1 Test case for out-of-synch handling in the UE for a definition of points A - F.
Time between points A and B. The test set's DPCCH_Ec/Ior is set to
DPCCH Quality Ratio Qa-b
for the time
Segment Ta-b Interval
.
Time between points C and D. The test set's DPCCH_Ec/Ior is set to
DPCCH Quality Ratio Qb-d
for 200 ms plus
Segment Tc-d Interval
(because Tb-c = Toff = 200 ms).
Time between points D and E. The test set's DPCCH_Ec/Ior is set to
DPCCH Quality Ratio Qd-e
for the time
Segment Td-e Interval
.
DPCCH_Ec/Ior between points A and B. (Note: the DPCCH quality ratio before point A is determined by the value that
Cell 1 Connected DPCH Level
is set to before the measurement is initiated, see
Downlink Channel Levels
.)
The
Out-of-Synch Handling of Output Power
measurement does not return any measurement results at this time. Initiating the measurement only causes the test set to change its DPCCH_Ec/Ior as specified by the
Out-of-Synch Handling of Output Power Measurement Parameters
. You must use an external spectrum analyzer to measure the UE's resulting output power. When the
Out-of-Synch Handling of Output Power
measurement is initiated from the test set, the test set generates a trigger at point A that you can route to the test set's rear panel
TRIG OUT
connector using the
Trigger Output Setup
setting, to trigger the spectrum analyzer. See
How Do I Make an Out-of-Synch Handling of Output Power Measurement?
.