SETup:WCTFormat

SETup:WCTFormat

Last updated: January 16, 2009

This section is only applicable to the lab application.

GPIB Command Conventions

SETup:WCTFormat:COUNt[:SNUMber]

Function

The setting form of this command performs two functions: (1) sets the number of change of TFC measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe command to 1 (on).

The query form returns the number of change of TFC multi-measurements to be made when multi-measurement state is ON.

See Multi-Measurement Count .

Setting

Range: 1 to 999

Query

Range: 1 to 999

*RST Setting

Off (10 when STATe = On)

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:COUNt 5" 
! Turns on multi-measurements and sets the number of measurements to 5.

SETup:WCTFormat:COUNt:NUMBer

Function

Sets/queries the number of change of TFC measurements to be made when the multi-measurement state is ON. See Multi-Measurement Count .

Setting

Range: 1 to 999

Query

Range: 1 to 999

*RST Setting

10

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:COUNt:NUMBer 5" 
! Sets the number of change of TFC multi-measurements to 5.

SETup:WCTFormat:COUNt:STATe

Function

Sets/queries the change of TFC multi-measurement state. See Multi-Measurement Count .

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (off)

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:COUNt:STATe ON"
! Turns on change of TFC multi-measurements.

SETup:WCTFormat:RMCHannel<12|33|64|144|384>:STEP:DOWN[:SIZE]

Function

Sets/queries the Step Down Size .

Setting

Range: -15.0 to 0.0 dB

Resolution: 0.1 dB

Query

Same as setting.

*RST Setting

  • RMCHannel12: -7.0 dB
  • RMCHannel33: -7.0 dB
  • RMCHannel64: -10.0 dB
  • RMCHannel144: -12.0 dB
  • RMCHannel384: -12.0 dB

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:RMCHannel12:STEP:DOWN -8" 
! Sets the 12.2k RMC Step Down Size to -8 dB.
 
OUTPUT 714;"SETup:WCTFormat:RMCHannel384:STEP:DOWN -13"
! Sets the 384k RMC Step Down Size to -13 dB.

SETup:WCTFormat:RMCHannel<12|33|64|144|384>:STEP:UP[:SIZE]

Function

Sets/queries the Step Up Size .

Setting

Range: 0.0 to 15.0 dB

Resolution: 0.1 dB

Query

Same as setting.

*RST Setting

  • RMCHannel12: 7.0 dB
  • RMCHannel33: 7.0 dB
  • RMCHannel64: 10.0 dB
  • RMCHannel144: 12.0 dB
  • RMCHannel384: 12.0 dB

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:RMCHannel12:STEP:UP 8" 
! Sets the 12.2k RMC Step Up Size to 8 dB.
 
OUTPUT 714;"SETup:WCTFormat:RMCHannel384:STEP:UP 13"
! Sets the 384k RMC Step Up Size to 13 dB.

SETup:WCTFormat:RMCHannel<12|33|64|144|384>:STEP:ERRor:LIMit:LOWer

Function

Sets/queries the Step Error Limit: Lower .

Setting

Range: -15.0 to 0.0 dB

Resolution: 0.1 dB

Query

Same as setting.

*RST Setting

  • RMCHannel12: -2.0 dB
  • RMCHannel33: -2.0 dB
  • RMCHannel64: -2.0 dB
  • RMCHannel144: -3.0 dB
  • RMCHannel384: -3.0 dB

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:RMCHannel12:STEP:ERRor:LIMit:LOWer -3" 
! Sets the 12.2k RMC Step Error Limit: Lower to -3 dB.
 
OUTPUT 714;"SETup:WCTFormat:RMCHannel384:STEP:ERRor:LIMit:LOWer -4"
! Sets the 384k RMC Step Error Limit: Lower to -4 dB.

SETup:WCTFormat:RMCHannel<12|33|64|144|384>:STEP:ERRor:LIMit:UPPer

Function

Sets/queries the Step Error Limit: Upper .

Setting

Range: 0.0 to 15.0 dB

Resolution: 0.1 dB

Query

Same as setting.

*RST Setting

  • RMCHannel12: 2.0 dB
  • RMCHannel33: 2.0 dB
  • RMCHannel64: 2.0 dB
  • RMCHannel144: 3.0 dB
  • RMCHannel384: 3.0 dB

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:RMCHannel12:STEP:ERRor:LIMit:UPPer 3" 
! Sets the 12.2k RMC Step Error Limit: Upper to 3 dB.
 
OUTPUT 714;"SETup:WCTFormat:RMCHannel384:STEP:ERRor:LIMit:UPPer 4"
! Sets the 384k RMC Step Error Limit: Upper to 4 dB.

SETup:WCTFormat:TIMeout[:STIMe]

Function

The setting form of this command performs two functions: (1) sets the change of TFC measurement's timeout value, and (2) sets the change of TFC measurement's timeout STATe to 1 (on).

The query form returns the change of TFC measurement's timeout value.

See Measurement Timeout .

Setting

Range: 0.1 to 999.9 seconds

Resolution: 0.1 second

Query

Same as setting

*RST Setting

10 seconds, (off)

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:TIMeout 5" 
! Sets timeout state to ON and sets the timeout time to 5 seconds.

SETup:WCTFormat:TIMeout:STATe

Function

Sets/queries the change of TFC measurement's timeout state. See Measurement Timeout .

Setting

Range: 1|ON|0|OFF

Query

Range: 1|0

*RST Setting

0 (off)

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:TIMeout:STATe ON"
! Sets the change of TFC measurement's timeout state to ON.

SETup:WCTFormat:TIMeout:TIME

Function

Sets/queries the change of TFC measurement's timeout setting to be used when the timeout state is set to ON. See Measurement Timeout .

Setting

Range: 0.1 to 999.9 seconds

Resolution: 0.1 second

Query

Same as setting

*RST Setting

10 seconds

Requirements

Lab Application Revision: C.01 and above

Programming Example

 
OUTPUT 714;"SETup:WCTFormat:TIMeout:TIME 5"
! Sets the change of TFC measurement's timeout to five seconds.