[:SENSe]:CORRection:COLLect:METHod:SOLR <p1>,<p2>

(Write-Only) Sets the Cal Method to Short-Open-Load-Reciprocal Thru. (Also known as 2-port Unknown Thru).

This Cal Method is NOT available on the N9912A.

Prompt for, then measure standards:

  • OPEN, SHORt, and LOAD on the <p1> and <p2> source ports.

  • THRU connection between the ports.

The Unknown Thru Standard:

  • Can have up to about 40 dB of loss and long electrical length.

  • Must be reciprocal: S21=S12.

Learn more about FieldFox Cal Methods using SCPI.

Relevant Modes

CAT, NA

To Cal in VVM mode, perform a Cal in CAT or NA mode, then switch to VVM mode.

Parameters

 

<p1>,<p2>

Port numbers to be calibrated. Choose 1,2

Examples

CORR:COLL:METH:SOLR 1,2

See Cal examples

Query Syntax

Not Applicable

Default

Not Applicable