N5264B Measurement Receiver


The N5264B is a PNA-X with no Source Ports or Tunable receivers. This makes the N5264B a very fast and very sensitive IF receiver that has been designed specifically for antenna and radar cross-section (RCS) measurements. When used with external sources, the N5264B can make frequency-scan measurements of antennas, or make RCS measurements in time domain.

Typical N5264B Application

Far Field Outdoor Antenna Measurements


N5264B Options and Limitations

The N5264B has the following options:

Limitations

The following VNA features are NOT available on the N5264B

Configuring the N5264B

IF Receivers

Internal LO Source (Option 108) - 10 MHz to 26.5 GHz

The LO source output connector is on the N5264B rear panel.

The power level of the internal LO source is typically about +10 dBm and can NOT be changed.

To change the frequency of the LO Source:

  1. Change the IF frequency of the measurement if necessary.

  2. On the FOM dialog, change Receivers frequencies to the RF frequency range to be measured.

  3. The LO Source frequency range is set automatically. This frequency value can not be viewed.

External Sources  

Because the N5264B has no internal sources (except for the optional LO), newly created measurements are displayed with a source port = 0 (zero).

External sources can be configured so that they are controlled by the N5264B.  Learn how to configure an external source.

Once configured, all existing and new measurements are changed to use the external source as the source port. When two or more sources are configured, the first configured source is displayed by default.

The external source settings can be changed from the following dialogs:

How to Create Measurements

Create ratioed and unratioed measurements using the standard Receivers tab.

S-parameters are not available.

Calibration

The only calibrations available are Response, Source Power and Receiver calibrations.

These are performed from the Unguided Cal Wizard.