:CALCulate:EQUalization:CTLE:POLE1

Type

Command/Query

Syntax

:CALCulate:EQUalization:CTLE:POLE1 <numeric>

:CALCulate:EQUalization:CTLE:POLE1?

Description

This command sets the equalization CTLE (Continuous Time Linear Equalization) Pole1 parameter.

Parameter

Description

CTLE DC pole1 parameter

Data Type

Double precision floating point type (Double)

Range

0 - 2E10

Preset Value

1.95E9

Unit

Hz

Resolution

-

Query Response

{numeric}<newline><^END>

Examples

10 OUTPUT 717;":CALC:EQU:CTLE:POLE1 2.05E9"
20 OUTPUT 717;":CALC:EQU:CTLE:POLE1?"
30 ENTER 717;A$

Related Commands

:CALCulate:EQUalization:CTLE:DC

:CALCulate:EQUalization:CTLE:POLE2

:CALCulate:EQUalization:CTLE:ZERO1

:CALCulate:EQUalization:FILename

:CALCulate:EQUalization:STATe

:CALCulate:EQUalization:TYPE

Equivalent GUI

Setup > Adv Waveform > Equalization > Pole1 Freq  

Eye/Mask > Advanced Waveform (Under Stimulus) > Equalization > Pole1 Freq