SYSTem: COMMunicate Commands


Controls and queries settings that affect the VNA system.

SYSTem:COMMunicate:

ECAL

    | CATalog?

    | CLISt?

    | COUNt?

    | DMEMory

        | CLEar

        | IMPort

    | EXPort

        | SNP

    | INFormation?

    | KNAMe:INFormation?

    | LIST?

    | PATH:COUNt?

GPIB

    | PMETer

    | ADDRess

    | RDEVice

           | CLOSe

           | OPEN

           | READ?

           | RESet

           | WBINary

           | WBLock

           | WRITe

LAN:HOSTname

PSENsor

TCPip:CONTrol?

USB:PMETer:CAT?

VISA

    | RDEVice

           | CLOSe

           | FIND?

           | OPEN

           | READ?

           | RESet

           | TIMeout

           | WBINary

           | WRITe

Click on a keyword to view the command details.

See Also


SYSTem:COMMunicate:ECAL:CATalog?

Applicable Models: All

(Read-only) Returns the ID string of ECals that are connected to the VNA USB. Use the list to select a Ecal for Ecal calibration.

Parameters

 

Examples

SYST:COMM:ECAL:CAT?

system:communicate:ecal:catalog?

Return Type

Comma-delimited strings.

Default

Not applicable


SYSTem:COMMunicate:ECAL<mod>:CLISt?

Applicable Models: All

(Read-only) Returns a list of characterizations stored in the specified ECal module.

Parameters

 

<mod>

ECal module from which to read user characterization numbers. Choose from 1 to 50. If unspecified, value is set to 1.

Examples

Module 1 contains User Characterizations 1 and 3.

SYST:COMM:ECAL:CLIST?

'Returns the following (0 always indicates the factory characterization):

0,1,3

Return Type

Numeric list, separated by commas.

Default

Not Applicable


SYSTem:COMMunicate:ECAL:COUNt?

Applicable Models: All

(Read-only) Returns the number of installed cal kits.

Examples

SYST:COMM:ECAL:COUNt?

Query Syntax

SYST:COMM:ECAL:COUNt?

Return Type

Numeric

Default

Not Applicable


SYSTem:COMMunicate:ECAL:DMEMory:CLEar <kitName>

Applicable Models: All

(Write-only) Deletes user characterizations from VNA disk memory.

Parameters

 

<kitName>

Optional String argument.  ECal Model, User Characterization name + " ECal", and serial number of the ECal module, separated by spaces. See examples below.

If unspecified, ALL User Characterizations that are stored in VNA disk memory are deleted.

Examples

'These examples all use "MyUserChar" as the User characterization name.

'The "My User Char" characterization is deleted from disk memory.

SYST:COMM:ECAL:DMEM:CLE "N4433A MyUserChar ECal 00001"

'All User characterizations are deleted from disk memory.

SYST:COMM:ECAL:DMEM:CLE

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:ECAL:DMEMory:IMPort <file>

Applicable Models:All

(Write-only) After the VNA disk memory is Exported to a file, use this command to Import the file into VNA disk memory, which allows the User Characterization to be used with the VNA and ECal module.

Note: An ECal confidence check can NOT be performed remotely from User Characterizations that are stored on the VNA disk.

Parameters

 

<file>

String. Full path and file name of file that was exported.

Examples

SYST:COMM:ECAL:DMEM:IMP "c:\users\public\network analyzer\ECal User Characterizations/myDiskUserChar.euc"

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:ECAL:EXPort <kit>[,<file>][,<NewName>]]

Applicable Models: All

(Write-only) Saves an existing ECal characterization to a file. Use this command to archive the user characterization or to move the characterization to a different VNA for use with the specified ECal module. After exporting the user characterization, use SYST:COMM:ECAL:DMEM:IMPort to make the user characterization available for use.

Parameters

 

<kit>

String. Not case sensitive. ECal Model, User char name + " ECal", and serial number of the ECal module used for the characterization, separated by spaces. See examples below.

If the model and serial number of the module is not found, an error is returned.

[<file>]

Optional String argument. Path and filename of the user characterization. If not specified, the file is saved using characterization name + ".euc". If the path is not specified, it is stored in C:/Program Files/Keysight/Network Analyzer/ECal User Characterizations/. The extension ".euc" is appended if one is not specified.

[<NewName>]

Optional String argument. This allows you to change the name for the User Characterization. When specified, the new name is saved in the file with the characterization. If unspecified, the existing user characterization name is saved.

Note: If this argument is specified, the second argument (<file>) must also be specified.

Examples

'These examples all use "MyUserChar" as the User characterization name.

'All parameters specified

SYST:COMM:ECAL:EXP "N4433A MyUserChar ECal 00001","myUserChar.euc","NewUserChar"

'First two parameters are specified

system:communicate:ecal:export "N4691B MyUserChar ECal 00500","myUserChar.euc"

'Only first parameter is specified

SYST:COMM:ECAL:EXP "N4433A MyUserChar ECal 00001"

Query Syntax

Not Applicable

Default

Not Applicable

 


SYSTem:COMMunicate:ECAL:EXPort:SNP <kit>,<ecalState>,<snpFileName>

Applicable Models: All

(Write-only) Read S parameter of ECal Thru from the ECal memory and save it as s2p file.

Parameters

 

<kit>

String. Not case sensitive. ECal Model, User char name + " ECal", and serial number of the ECal module used for the characterization, separated by spaces. See examples below.

If the model and serial number of the module is not found, an error is returned.

<ecalState>

ECal transmission path.  Choose from AB, AC, AD, BA, BC, BD, CA, CB, CD, DA, DB or DC. Not case sensitive.

<snpFileName>

Path and filename of the output s2p file name.

Examples

SYST:COMM:ECAL:EXP "N4433A ECal 00001","BC","D:\ecalthru.s2p"

Query Syntax

Not Applicable

Default

Not Applicable

 


SYSTem:COMMunicate:ECAL<mod>:INFormation? [<char>]

Applicable Models: All

(Read-only)  Reads the identification and characterization information from the specified ECal module.

Note: To read user-characterization information that is stored in VNA disk memory, then use SYST:COMM:ECAL:KNAM:INF?

Parameters

 

<mod>

ECal module from which to read characterizations. Choose from 1 through 50. If unspecified, value is set to 1.

Do NOT assume the <mod> number is the order in which ECal modules were connected.

Use SYST:COMM:ECAL:LIST? to read a list of <mod> numbers of currently-attached ECal modules.

<char>

Optional argument. Specifies which characterization to read information from.  If not specified, value is set to CHAR0.

Choose from:

  • CHAR0   Factory characterization (data that was stored in the ECal module by Keysight)

  • CHAR1   User characterization #1

  • CHAR2   User characterization #2

  •  - through -

  • CHAR12   User characterization #12

Examples

SYST:COMM:ECAL2:INFormation? char5

'Example return string:

"ModelNumber: 85092-60007, SerialNumber: 01386, ConnectorType: N5FN5F, PortAConnector: Type N (50) female, PortBConnector: Type N (50) female, MinFreq: 30000, MaxFreq: 9100000000, NumberOfPoints: 250, Calibrated: July 4 2002"

Return Type

Character

Default

Not Applicable


SYSTem:COMMunicate:ECAL:KNAMe:INFormation? <kitName>

Applicable Models: All

(Read-only) Reads the identification and characterization information from the specified ECal module or VNA disk memory.

Learn more about User Characterization in VNA Disk Memory.

Parameters

 

<kitName>

String. ECal model and characterization to read information from, enclosed in quotes, in the following format:

<model> <name> ECal <serial number>

Where:

<model>: Always required

<name>:

  • For the factory characterization, do not specify.

  • For a user-characterization stored in the module, use User <n> in the string, where <n> is the user-characterization number. Not case sensitive. Separate User and <n> with a space.

  • For a user-characterization stored in VNA disk memory, use <charName> from SENS:CORR:CKIT:ECAL:CHAR:DMEM:SAVE <charName>

ECal - not case sensitive

<serial number>: Optional. Include when two or more ECal modules with same model number are attached to the VNA,

Each item is separated with a space.

Examples

'For a factory characterization in module memory:
SYST:COMM:ECAL:KNAM:INF? "N4433A ECal"

'For user characterization in module memory with optional serial number:
SYST:COMM:ECAL:KNAM:INF? "N4433A User 1 ECal 00028"

'For user characterization "foo" in disk memory:
SYST:COMM:ECAL:KNAM:INF? "N4433A foo ECal 00028"

'Example return string:

"ModelNumber: N4433A, SerialNumber: 00028, ConnectorType: N5FN5F, PortAConnector: Type N (50) female, PortBConnector: Type N (50) female, MinFreq: 30000, MaxFreq: 9100000000, NumberOfPoints: 250, Calibrated: July 4 2002"

Return Type

String

Default

Not Applicable


SYSTem:COMMunicate:ECAL:LIST?

Applicable Models: All

(Read-only) Returns a list of index numbers for ECal modules that are currently attached to the VNA. Use these numbers (called <mod> in VNAHelp) to refer to the ECal module using SCPI commands.

Examples

SYST:COMM:ECAL:LIST?

'If 2 modules are attached to the VNA
'then the returned list will be:

+1,+2

'If NO modules are attached to the VNA
'then the returned list will be:

+0

Return Type

Numeric list, separated by commas.

Default

Not Applicable


SYSTem:COMMunicate:ECAL<n>:PATH:COUNt? <path>

Applicable Models: All

(Read-only)  Returns the number of unique states that exist for the specified path name on the selected ECal module.

This command performs exactly the same function as CONT:ECAL:MOD:PATH:COUNt?

Use the CONT:ECAL:MOD:PATH:STAT command to set the module into one of those states.  

Use SENS:CORR:CKIT:ECAL:PATH:DATA? to read the data for a state.

Parameters

 

<n>

USB number of the ECal module.  Choose from 1 to 50.

If unspecified (only one ECal module is connected to the USB), <n> is set to 1. If two or more modules are connected, use SYST:COMM:ECAL:LIST? to determine how many, and SYST:COMM:ECAL:INF? to verify their identities.

<path>

Name of the path for which to read number of states.  Choose from:

Reflection paths

  • A

  • B

  • C (4-port modules)

  • D (4-port modules)

Transmission paths

  • AB

  • AC (4-port modules)

  • AD (4-port modules)

  • BC (4-port modules)

  • BD (4-port modules)

  • CD (4-port modules)

Examples

SYST:COMM:ECAL:PATH:COUNt?
system:communicate:ecal:path:count?

Return Type

Integer

Default

Not Applicable


SYSTem:COMMunicate:GPIB:PMETer[:ADDRess] <num>  Superseded

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

Note: This command is replaced with SYST:COMM:PSENsor

(Read-Write) Specifies the GPIB address of the power meter to be used in a source power calibration. When performing a source power cal, the VNA will search VISA interfaces that are configured in the Keysight IO LIbraries on the VNA.

Parameters

 

<num>

GPIB address of the power meter. Choose any integer between 0 and 30.

Examples

SYST:COMM:GPIB:PMET 13

system:communicate:gpib:pmeter:address 14

Query Syntax

SYSTem:COMMunicate:GPIB:PMETer[:ADDRess]?

Return Type

Numeric

Default

13


SYSTem:COMMunicate:GPIB:RDEVice:CLOSe <ID>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Write only) Closes the remote GPIB session. This command should be sent when ending every successful OPEN session.

Parameters

 

<ID>

Session identification number that was returned with the OPEN? command.

Examples

See an example program

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:OPEN <bus>, <addr>, <timeout>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Read-Write) Initiates a GPIB pass-through session. First send this OPEN command, then send the OPEN query to read the session ID number. An existing GPIB pass-through session remains open after an instrument preset.

To learn more about GPIB pass-through capability, see the example program.

Parameters

 

<bus>

Bus ID number.

You can find the USB-GPIB adapter bus number by looking at the dialog that appears when the USB-GPIB device is connected. Error 1073 indicates the bus or address number is incorrect.

Use 0 (zero) when connected using a GPIB cable to the VNA controller port.

<addr>

GPIB Address of the device to be controlled

<timeout>

The amount of time (in milliseconds) to wait for a response from the remote device after sending a command.  A "timeout" error is displayed after this time has passed without a response.

Examples

See an example program

Query Syntax

SYSTem:COMMunicate:GPIB:RDEVice:OPEN?

Returns the session identification number that is used when communicating with this device.

Return Type

Numeric

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:READ? <ID>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Read-only) Returns data from the GPIB pass-through device.

Parameters

 

<ID>

Session identification number that was returned with the OPEN? command.

Examples

See an example program

Return Type

String

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:RESet

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Write-only)   Performs the same function as SYST:COMM:GPIB:RDEV:CLOS except that ALL pass-through sessions are closed.

Examples

SYST:COMM:GPIB:RDEV:RES

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:WBINary <ID>,<data>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Write-only)  Sends data to a GPIB pass-through device. This command requires a header that specifies the size of the data to be written. The header (described below) is not passed along to the device.

Use this command if too many embedded quotes prevent you from using SYST:COMM:GPIB:RDEV:WRIT.

Use SYST:COMM:GPIB:RDEV:OPEN to open the pass through session.

Parameters

 

<ID>

Session identification number that was returned with the OPEN? command.

<data>

Data to be sent to the GPIB pass-through device. Use the following syntax:

#<num digits><byte count><data bytes><NL><END>

<num_digits> specifies how many digits are contained in <byte_count>

<byte_count> specifies how many data bytes will follow in <data bytes>

Examples

SYSTem:COMMunicate:GPIB:RDEVice:WBINary 101,#17ABC+XYZ<nl><end>

# - always sent before data.

1 - specifies that the byte count is one digit (7).

7 - specifies the number of data bytes that will follow, not counting <NL><END>.

ABC+XYZ - Data block

<nl><end> - always sent at the end of block data.

 

The following example sends a line feed at the end.

SYST:COMM:GPIB:RDEV:WBIN 1,#210SYST:PRES<EOL>

The <EOL> represents your linefeed character.

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:WBLock <ID>,<data>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Write-only)  Same as SYSTem:COMM:GPIB:RDEV:WBIN (above) but the header IS passed along to the device.

Use this command if too many embedded quotes prevent you from using SYST:COMM:GPIB:RDEV:WRIT.

Parameters

 

<ID>

Session identification number that was returned with the OPEN? command.

<data>

Data to be sent to the GPIB pass-through device. See previous command.

Examples

See previous example.

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:GPIB:RDEVice:WRITe <ID>,<string>

Applicable Models: N522xB, N523xB, N524xB, M937xA, P937xA

(Write-only) Sends ASCII string data to the GPIB pass-through device.

A line feed is NOT appended to the string data. To send a line feed, see the example in SYST:COMM:GPIB:RDEV:WBIN.

Parameters

 

<ID>

Session identification number that was returned with the OPEN? command.

<string>

Commands to be sent to the GPIB pass-through device.

Examples

See an example program

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:LAN:HOSTname?

Applicable Models: All

(Read-only) Returns the LAN hostname that is visible in the Help, About Network Analyzer dialog box. Learn more. This is the same information that is visible on the LXI compliance dialog.

Parameters

None

Example

SYST:COMM:LAN:HOSTname?

Return Type

String

Default

Not applicable


SYSTem:COMMunicate:PSENsor <char>, <string>

Applicable Models: All

This command replaces SYST:COMM:GPIB:PMET:ADDR.

(Read-Write) Specifies the type and location of the power meter to be used in a source power calibration.

Parameters

 

<char>

Type of power meter/ sensor. Choose from:

  • GPIB  GPIB power meter

  • USB  USB power sensor or USB power sensor

  • LAN  LAN enabled power meter

  • ANY  Any VISA resource string or a visa alias

<string>

For GPIB, address of the power meter. Choose any integer between 0 and 30.

For USB, the ID string of the power meter or power sensor. Use SYST:COMM:USB:PMET:CAT? to see a list of ID strings of connected power meters and sensors.

For LAN, the hostname or IP address of the power meter.

For ANY,  any VISA resource string or a visa alias.

Examples

SYST:COMM:PSEN gpib, "14"

system:communicate:psensor usb, "Keysight Technologies,U2000A,MY12345678"

syst:comm:psen lan, "mymeter.Keysight.com"

syst:comm:psen any, "TCPIP0::mymeter.Keysight.com::5025::SOCKET"

Query Syntax

SYSTem:COMMunicate:PSENsor?

Return Type

Character / String

Default

GPIB


SYSTem:COMMunicate:USB:PMETer:CATalog?

Applicable Models: All

(Read-only) Returns the ID string of power meters / sensors that are connected to the VNA USB. Use the list to select a power sensor for a source power cal.

These meter/sensor ID strings can NOT be used as the resource string for configuring a USB-based PMAR (SYST:CONF:EDEV:IOConfig).

Parameters

 

Examples

SYST:COMM:USB:PMET:CAT?

system:communicate:usb:pmeter:catalog?

Return Type

Comma-delimited strings. Two power sensor strings are separate by a semicolon.

Default

Not applicable


SYSTem:COMMunicate:TCPip:CONTrol?

Applicable Models: All

(Read-only) Queries the TCP/IP port number to use for opening a TCP/IP socket control connection to the VNA.  The control connection is used for two purposes:

  1. To perform a Device Clear operation on the VNA

  2. To detect when a Service Request (SRQ) event occurs on the VNA.

The port number can range from 5000 to 5099. The VNA will skip over 5025 as it is being used for the primary socket connection.

To detect an SRQ, your program sends the appropriate commands via the regular socket connection to set up for a SRQ event to occur the same sequence of commands as if you were sending them via GPIB.  You write your program so that while your program is doing SCPI transactions on the standard socket connection, a second thread of execution in your program detects the SRQ on the control connection and responds to the event.  When the SRQ event occurs, the VNA sends a SRQ +xxx/n message on the control connection (where /n is linefeed character, ASCII value 10 decimal).  The xxx value in the SRQ +xxx/n string is the IEEE 488.2 status byte at the time the SRQ was generated.  So listening for that on the control connection is how your program detects the event.  If for your socket communication youre using a software API that provides for asynchronous communication via a callback mechanism (for example, if youre using Microsofts winsock API, or their .NET Socket class as in the example program below), in that case your listener execution thread is created implicitly for you so your program doesnt have to create one explicitly.

Note: If this SCPI query is sent to the VNA via a SCPI parser other than a TCP/IP socket connection (for example, if sent via GPIB), the query is not applicable in that case and will return value of 0.

Parameters

None

Example

See example program

Return Type

Integer

Default

Not applicable


SYSTem:COMMunicate:VISA:RDEVice:CLOSe <ID>

Applicable Models: N522xB, N523xB, N524xB, E5080A

(Write only) Closes the specified remote VISA session. VISA sessions should always be closed when you are finished communicating with the remote device. Use this command to close (end) each VISA session that was opened successfully using the OPEN command. If you have more than one open session, and need to close them all at the same time, it may be faster and easier to use the RESet command.

Parameters

 

<ID>

VISA session number (see  SYST:COMM:VISA:RDEV:OPEN[?]).

Examples

SYST:COMM:VISA:RDEV:CLOS 1

system:communicate:visa:rdevice:close 2

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:VISA:RDEVice:FIND? <VISA regex> [,<ADDRess|ALIas>]

Applicable Models: All

(Read-only) Returns a comma separated list of either VISA address strings or aliases.

Parameters

 

<VISA regex>

(String) VISA regular expressions are expressions defined by the user to find devices that have been set up on the VISA interface. The following are examples of VISA regular expressions:

Interface

Expression

GPIB

GPIB[0-9]*::?*INSTR

PXI

PXI?*INSTR

VXI

VXI?*INSTR

GPIB-VXI

GPIB-VXI?*INSTR

GPIB and GPIB-VXI

GPIB?*INSTR

All VXI

?*VXI[0-9]*::?*INSTR

ASRL

ASRL[0-9]*::?*INSTR

All

?*INSTR  or  ?*

Note that using "INSTR" in the VISA regular expression finds "instruments." To search all interfaces, use "?*".

<ADDRess|ALIas>

Optional. Determines whether addresses or aliases are returned.

Note: The list of aliases may have less‒or more‒entries than the list of addresses because not all addresses will have aliases, and one address can have more than one alias.

Examples

SYST:COMM:VISA:RDEV:FIND? "?*",ADDR

system:communicate:visa:rdevice:find? '?*INSTR',alias

Return Type

Variant

Default

Addresses returned if no return-type specified


SYSTem:COMMunicate:VISA:RDEVice:OPEN <addr>, <timeout>

Applicable Models: N522xB, N523xB, N524xB, E5080A

(Read-Write) Initiates a VISA pass-through session for a device. Immediately after successfully sending this command, send an OPEN? query to retrieve the unique session ID that is to be used whenever communicating with the device. Pass-through sessions can be closed by using the CLOSe command, the RESet command, or by properly shutting down the instrument or the analyzer application. Presetting the instrument will not close existing pass-through sessions.

Note: When opening a socket session (addresses of type: “TCPIP[board]::host address::port::SOCKET”) , you must use the appropriate VISA Address for the identifier argument. Using an alias to open a socket session is not currently supported. Aliases are allowed for all other types of supported sessions.

Parameters

 

<addr>

VISA Address or alias name of the device to be controlled

<timeout>

The amount of time (in milliseconds) to wait for a response from the remote device after sending a command.  A "timeout" error is displayed after this time has passed without a response.

Examples

SYST:COMM:VISA:RDEV:OPEN 'TCPIP0::A-N5242A-10096::hislip1::INSTR',1000

system:communicate:visa:rdevice:open 'MyAliasName',7000

Query Syntax

SYSTem:COMMunicate:VISA:RDEVice:OPEN?

Returns the VISA session identification number to be used when communicating with this device.

Return Type

Numeric

Default

Not Applicable


Applicable Models: N522xB, N523xB, N524xB, E5080A

(Read-only) Returns data from the VISA pass-through device.

Parameters

 

<ID>

VISA session number (see  SYST:COMM:VISA:RDEV:OPEN[?]).

Examples

SYST:COMM:VISA:RDEV:READ? 1

system:communicate:visa:rdevice:read? 3

Return Type

String

Default

Not Applicable


SYSTem:COMMunicate:VISA:RDEVice:RESet

Applicable Models: N522xB, N523xB, N524xB, E5080A

(Write-only)   Closes all currently open VISA pass-through sessions. See also CLOSe to close only one session at a time.

Examples

SYST:COMM:VISA:RDEV:RES

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:VISA:RDEVice:TIMeout <ID>, <timeout>

Applicable Models: All

(Read-Write) Sets or returns the timeout value (in milliseconds) for VISA pass-through commands for the specified VISA session ID.

Parameters

 

<ID>

VISA session number that was returned with the OPEN? command.

<timeout>

The amount of time (in milliseconds) to wait for a response from the remote device after sending a command. A "timeout" error is displayed after this time has passed without a response.

Examples

SYST:COMM:VISA:RDEV:TIM 1,6000

system:communicate:visa:rdevice:timeout 3,6000

Query Syntax

SYSTem:COMMunicate:VISA:RDEVice:TIMeout? <ID>

Returns the timeout value for the specified session ID.

Return Type

Numeric

Default

2000


SYSTem:COMMunicate:VISA:RDEVice:WBINary <ID>,<data>

Applicable Models: N522xB, N523xB, N524xB, E5080A

(Write-only)  Sends data to a VISA pass-through device. This command requires a header that specifies the size of the data to be written. The header (described below) is not passed along to the device.

Parameters

 

<ID>

VISA session number (see  SYST:COMM:VISA:RDEV:OPEN[?]).

<data>

Data to be sent to the VISA pass-through device. Use the following syntax:

#<num digits><byte count><data bytes><NL><END>

<num_digits> specifies how many digits are contained in <byte_count>

<byte_count> specifies how many data bytes will follow in <data bytes>

Examples

SYSTem:COMMunicate:VISA:RDEVice:WBINary 1,#17ABC+XYZ<nl><end>

# - always sent before data.

1 - specifies that the byte count is one digit (7).

7 - specifies the number of data bytes that will follow, not counting <NL><END>.

ABC+XYZ - Data block

<nl><end> - always sent at the end of block data.

 

The following example sends a line feed at the end.

SYST:COMM:VISA:RDEV:WBIN 1,#210SYST:PRES<EOL>

The <EOL> represents your linefeed character.

Query Syntax

Not Applicable

Default

Not Applicable


SYSTem:COMMunicate:VISA:RDEVice:WRITe <ID>,<string>

Applicable Models: N522xB, N523xB, N524xB, E5080A

(Write-only) Sends ASCII string data to the VISA pass-through device. If sending a command that returns data, follow with the READ? query.

Parameters

 

<ID>

VISA session number (see  SYST:COMM:VISA:RDEV:OPEN[?]).

<string>

Commands to be sent to the VISA pass-through device.

Examples

SYST:COMM:VISA:RDEV:WRIT 1,'*IDN?'

system:communicate:visa:rdevice:write 2,'SYST:PRES'

Query Syntax

Not Applicable

Default

Not Applicable