Configures and makes settings for an external Keysight 81110A Pulse Generator.
SYST:CONF:EDEVice:PULSe | CHAN | HAMP | LAMP | LIMP | MMODe | SIMP |
Click on a keyword to view the command details.
Root SYST:CONF:EDEV commands
Learn about: Configure an External Pulse Generator
Learn about Configure and External Device
SYST:PREF:ITEM:EDEV:DPOL - Determines whether External Devices remain activated or are de-activated when the VNA is Preset or when a Instrument State is recalled.
Applicable Models: All (Read-Write) Sets and returns the output channel of the pulse generator. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<value> |
Pulse Generator output port. Choose from 1 or 2. |
Examples |
SYST:CONF:EDEV:PULS:CHAN "81110",1 system:configure:edevice:pulse:chan "myPG",2 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:CHAN? <name> |
Return Type |
Numeric |
1 |
Applicable Models: All (Read-Write) Sets and returns the High amplitude (voltage) of the pulse generator. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<value> |
Pulse Generator high amplitude voltage. |
Examples |
SYST:CONF:EDEV:PULS:HAMP "81110",3 system:configure:edevice:pulse:HAMP "myPG",4 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:HAMP? <name> |
Return Type |
Numeric |
5 |
Applicable Models: All (Read-Write) Pulse Generator low amplitude voltage. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<value> |
Pulse Generator low amplitude voltage. |
Examples |
SYST:CONF:EDEV:PULS:LAMP "81110",.2 system:configure:edevice:pulse:lamp "myPG",1 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:LAMP? <name> |
Return Type |
Numeric |
0 |
Applicable Models: All (Read-Write) Sets and returns the load impedance of the pulse generator. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<value> |
Pulse generator load impedance. |
Examples |
SYST:CONF:EDEV:PULS:LIMP "81110",52 system:configure:edevice:pulse:limp "myPG",49 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:LIMP? <name> |
Return Type |
Numeric |
50 |
Applicable Models: All (Read-Write) Sets and returns the Master (On/Off) setting of the external pulse generator. The ON setting allows the external pulse generator to set the master clock frequency for the other pulse generators. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<bool> |
Master setting. Choose from: ON or 1 - Use the external pulse generator becomes the master clock frequency. OFF or 0 - Use the internal pulse generator as the master clock frequency. |
Examples |
SYST:CONF:EDEV:PULS:MMOD "81110",OFF system:configure:edevice:pulse:mmode "myPG",1 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:MMODe? <name> |
Return Type |
Boolean |
OFF or 0 |
Applicable Models: All (Read-Write) Sets and returns the source impedance of the pulse generator. |
|
Parameters |
|
<name> |
String - Name of the external pulse generator. |
<value> |
Pulse generator source impedance. |
Examples |
SYST:CONF:EDEV:PULS:SIMP "81110",52 system:configure:edevice:pulse:simp "myPG",49 |
Query Syntax |
SYSTem:CONFigure:EDEVice:DC:PULSe:SIMP? <name> |
Return Type |
Numeric |
50 |