Last updated: January 16, 2009
This section is only applicable to the lab application.
Queries a change of TFC measurement result and returns the following comma separated values: Integrity Indicator, Step Down Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Down Error, Step Down Pass/Fail, Step Up Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Up Error, Step Up Pass/Fail. |
|
Integrity Indicator - See Integrity Indicator . |
|
OUTPUT 714;"FETCh:WCTFormat?" |
Queries the intermediate count of measurements completed in a multi-measurement. To set up multi-measurements, see SETup:WCTFormat:COUNt[:SNUMber] . |
|
OUTPUT 714;"FETCh:WCTFormat:ICOunt?" ! Returns the intermediate count of change of TFC multi-measurements. |
Returns the Integrity Indicator for a change of TFC measurement result. |
|
OUTPUT 714;"FETCh:WCTFormat:INTegrity?" |
Queries a change of TFC measurement result and returns the following Step Down Relative Power (in dB) result(s):
To set up multi-measurements, see SETup:WCTFormat:COUNt[:SNUMber] . |
|
OUTPUT 714;"FETCh:WCTFormat:STEP:DOWN:RELative:POWer?" ! Returns the Step Down Relative Power average result in dB. OUTPUT 714;"FETCh:WCTFormat:STEP:DOWN:RELative:POWer:MAXimum?" ! Returns the Step Down Relative Power maximum result in dB. |
Queries a change of TFC measurement result and returns the Step Down Error (in dB) result. |
|
OUTPUT 714;"FETCh:WCTFormat:STEP:DOWN:ERRor?" |
Queries a change of TFC measurement result and returns the Step Down Pass/Fail result. |
|
Range: 0 (pass), 1 (fail), 9.91E+37 | |
OUTPUT 714;"FETCh:WCTFormat:STEP:DOWN:FAIL?" |
Queries a change of TFC measurement result and returns the following Step Up Relative Power (in dB) result(s):
To set up multi-measurements, see SETup:WCTFormat:COUNt[:SNUMber] . |
|
OUTPUT 714;"FETCh:WCTFormat:STEP:UP:RELative:POWer?" ! Returns the Step Up Relative Power average result in dB. OUTPUT 714;"FETCh:WCTFormat:STEP:UP:RELative:POWer:MAXimum?" ! Returns the Step Up Relative Power maximum result in dB. |
Queries a change of TFC measurement result and returns the Step Up Error (in dB) result. |
|
OUTPUT 714;"FETCh:WCTFormat:STEP:UP:ERRor?" |
Queries a change of TFC measurement result and returns the Step Up Pass/Fail result. |
|
Range: 0 (pass), 1 (fail), 9.91E+37 | |
OUTPUT 714;"FETCh:WCTFormat:STEP:UP:FAIL?" |